Successful and reproducible myocardial opacification during two-dimensional echocardiography from right heart injection of contrast.

Abstract

BACKGROUND

Myocardial contrast echocardiography currently involves intro-arterial injection of contrast. For this technique to have a broader application, it is necessary that myocardial opacification be achieved from a venous injection of contrast.

METHODS AND RESULTS

To achieve myocardial opacification after right-side injection of contrast, two groups of open-chest anesthetized dogs were studied. Group 1 included nine dogs in whom microbubbles of various sizes, concentrations, and volumes were injected into the left atrium to determine microbubble characteristics that influence myocardial opacification. Group 2 included eight dogs in whom the effect of the combination of microbubble characteristics and myocardial blood flow on myocardial opacification was evaluated after right atrial injection of contrast. Background-subtracted time-intensity plots were generated from the myocardium to measure peak videointensity. In the group 2 dogs, digital subtraction and color coding were used to further highlight the contrast effect. The number, concentration, and size of the microbubbles all independently affected (p less than 0.01) peak myocardial videointensity after left atrial injection of contrast on multivariate analysis. Highly concentrated microbubbles (4.4 to 5.1 billion/ml) given during dipyridamole-induced coronary hyperemia was most frequently (88%) associated with myocardial opacification after right atrial injection of contrast and was the best predictor of this result on multivariate analysis (chi 2= 9.01, p = 0.003). No changes were noted in left atrial, left ventricular, and pulmonary artery pressures despite injection of large numbers of microbubbles into the right atrium.

CONCLUSIONS

Successful and reproducible myocardial opacification can be achieved during myocardial contrast echocardiography after right atrial injection of contrast. These findings could have far-reaching implications in the use of myocardial contrast echocardiography in acute and chronic ischemic syndromes in humans.

Formats available

You can view the full content in the following formats:

eLetters(0)

eLetters should relate to an article recently published in the journal and are not a forum for providing unpublished data. Comments are reviewed for appropriate use of tone and language. Comments are not peer-reviewed. Acceptable comments are posted to the journal website only. Comments are not published in an issue and are not indexed in PubMed. Comments should be no longer than 500 words and will only be posted online. References are limited to 10. Authors of the article cited in the comment will be invited to reply, as appropriate.

Comments and feedback on AHA/ASA Scientific Statements and Guidelines should be directed to the AHA/ASA Manuscript Oversight Committee via its Correspondence page.

Information & Authors

Information

Published In

Go to Circulation
Go to Circulation
Circulation
Pages: 1557 - 1564
PubMed: 1555293

History

Published online: 1 April 1992
Published in print: April 1992

Permissions

Request permissions for this article.

Authors

Affiliations

F S Villanueva
Department of Medicine, University of Virginia School of Medicine, Charlottesville.
W P Glasheen
Department of Medicine, University of Virginia School of Medicine, Charlottesville.
J Sklenar
Department of Medicine, University of Virginia School of Medicine, Charlottesville.
A R Jayaweera
Department of Medicine, University of Virginia School of Medicine, Charlottesville.
S Kaul
Department of Medicine, University of Virginia School of Medicine, Charlottesville.

Metrics & Citations

Metrics

Citations

Download Citations

If you have the appropriate software installed, you can download article citation data to the citation manager of your choice. Select your manager software from the list below and click Download.

  1. Microbubble Formulations: Synthesis, Stability, Modeling and Biomedical Applications, Ultrasound in Medicine & Biology, 45, 2, (301-343), (2019).https://doi.org/10.1016/j.ultrasmedbio.2018.09.022
    Crossref
  2. Testing the Efficacy of Contrast-Enhanced Ultrasound in Detecting Transplant Rejection Using a Murine Model of Heart Transplantation, American Journal of Transplantation, 17, 7, (1791-1801), (2017).https://doi.org/10.1111/ajt.14180
    Crossref
  3. Myocardial Contrast Echocardiography, JACC: Cardiovascular Imaging, 3, 2, (212-218), (2010).https://doi.org/10.1016/j.jcmg.2009.11.003
    Crossref
  4. Enhanced Left Ventricular Endocardial Border Delineation with an Intravenous Injection of SonoVue, a New Echocardiography Contrast Agent:, Echocardiography, 17, 8, (705-711), (2007).https://doi.org/10.1111/j.1540-8175.2000.tb01223.x
    Crossref
  5. Clinical Experience with SonoVue in Myocardial Perfusion Imaging, Echocardiography, 17, s6, (2007).https://doi.org/10.1111/j.1540-8175.2000.tb01190.x
    Crossref
  6. Imaging of Myocardial Perfusion with SonoVue™ in Patients with a Prior Myocardial Infarction, Echocardiography, 16, s1, (753-760), (2007).https://doi.org/10.1111/j.1540-8175.1999.tb00146.x
    Crossref
  7. Real‐Time Contrast Echo Assessment of Myocardial Perfusion at Low Emission Power, Echocardiography, 16, 8, (799-809), (2007).https://doi.org/10.1111/j.1540-8175.1999.tb00132.x
    Crossref
  8. Reproducibility of Myocardial Contrast Echocardiography in Human Studies, Echocardiography, 10, 3, (255-263), (2007).https://doi.org/10.1111/j.1540-8175.1993.tb00035.x
    Crossref
  9. INTRAVASCULAR ULTRASOUND CONTRAST MEDIA, Veterinary Radiology & Ultrasound, 36, 4, (307-314), (2005).https://doi.org/10.1111/j.1740-8261.1995.tb00267.x
    Crossref
  10. Pulmonary Transit of Echocontrast Agents during Mechanical Ventilation: A Clinical Transthoracic Echocardiographic Study, Echocardiography, 22, 5, (395-401), (2005).https://doi.org/10.1111/j.1540-8175.2005.04084.x
    Crossref
  11. See more
Loading...

View Options

View options

PDF and All Supplements

Download PDF and All Supplements

PDF/EPUB

View PDF/EPUB
Login options

Check if you have access through your login credentials or your institution to get full access on this article.

Personal login Institutional Login
Purchase Options

Purchase this article to access the full text.

Purchase access to this article for 24 hours

Successful and reproducible myocardial opacification during two-dimensional echocardiography from right heart injection of contrast.
Circulation
  • Vol. 85
  • No. 4

Purchase access to this journal for 24 hours

Circulation
  • Vol. 85
  • No. 4
Restore your content access

Enter your email address to restore your content access:

Note: This functionality works only for purchases done as a guest. If you already have an account, log in to access the content to which you are entitled.

Figures

Tables

Media

Share

Share

Share article link

Share

Comment Response