Efficacy and Safety of a Novel Bioabsorbable Polymer-Coated, Everolimus-Eluting Coronary Stent

The EVOLVE II Randomized Trial
Originally publishedhttps://doi.org/10.1161/CIRCINTERVENTIONS.114.002372Circulation: Cardiovascular Interventions. 2015;8:e002372

    Abstract

    Background—

    Drug eluting stents with durable polymers may be associated with hypersensitivity, delayed healing, and incomplete endothelialization, which may contribute to late/very late stent thrombosis and the need for prolonged dual antiplatelet therapy. Bioabsorbable polymers may facilitate stent healing, thus enhancing clinical safety. The SYNERGY stent is a thin-strut, platinum chromium metal alloy platform with an ultrathin bioabsorbable Poly(D,L-lactide-co-glycolide) abluminal everolimus-eluting polymer. We performed a multicenter, randomized controlled trial for regulatory approval to determine noninferiority of the SYNERGY stent to the durable polymer PROMUS Element Plus everolimus-eluting stent.

    Methods and Results—

    Patients (n=1684) scheduled to undergo percutaneous coronary intervention for non–ST-segment–elevation acute coronary syndrome or stable coronary artery disease were randomized to receive either the SYNERGY stent or the PROMUS Element Plus stent. The primary end point of 12-month target lesion failure was observed in 6.7% of SYNERGY and 6.5% PROMUS Element Plus treated subjects by intention-to-treat (P=0.83 for difference; P=0.0005 for noninferiority), and 6.4% in both the groups by per-protocol analysis (P=0.0003 for noninferiority). Clinically indicated revascularization of the target lesion or definite/probable stent thrombosis were observed in 2.6% versus 1.7% (P=0.21) and 0.4% versus 0.6% (P=0.50) of SYNERGY versus PROMUS Element Plus–treated subjects, respectively.

    Conclusions—

    In this randomized trial, the SYNERGY bioabsorbable polymer everolimus-eluting stent was noninferior to the PROMUS Element Plus everolimus-eluting stent with respect to 1-year target lesion failure. These data support the relative safety and efficacy of SYNERGY in a broad range of patients undergoing percutaneous coronary intervention.

    Clinical Trial Registration—

    URL: http://www.clinicaltrials.gov. Unique identifier: NCT01665053.

    WHAT IS KNOWN

    • Durable polymer on coronary drug-eluting stents may be associated with inflammation, neoatherosclerosis, and thrombosis.

    • Bioabsorbable polymers may facilitate healing and enhanced clinical safety.

    WHAT THE STUDY ADDS

    • In a prospective, multicenter randomized single-blind trial, the Synergy stent, with a bioabsorbable polymer, proved noninferior to the PROMUS Element Plus stent for target lesion failure to 1 year.

    • This study establishes comparable clinical safety and efficacy of everolimus elution from a bioabsorbable polymer thin strut platinum chromium metal platform in support of regulatory approval for this novel coronary stent device.

    Drug-eluting stents (DES) that deliver antiproliferative drugs from a durable polymer have significantly reduced restenosis compared with bare metal stents.1 However, durable polymers may be associated with inflammation, delayed healing, and incomplete endothelialization, which may contribute to the risk of late (30 days to 1 year) and very late (>1 year) stent thrombosis compared with bare metal stents.2 Whether metal alloy coronary stent platforms with bioresorbable polymers are associated with improved clinical outcomes when compared with newer durable polymer DES has been the subject of debate3,4 and may be influenced by additional factors, including stent strut thickness, polymer composition, distribution, and load.5 Although current American College of Cardiology/American Heart of Association (ACC/AHA) clinical practice guidelines recommend at least 12 months of dual antiplatelet therapy (DAPT) after DES deployment in patients who are not at increased risk for bleeding,1 recent studies suggest that even longer duration DAPT therapy (≥30 months) provides additional ischemic event reduction.6 The SYNERGY stent (Boston Scientific Corporation, Marlborough, MA) is a novel thin-strut platinum chromium (PtCr) metal alloy stent that elutes everolimus from an ultrathin bioabsorbable Poly(D,L-lactide-co-glycolide) polymer applied to the abluminal surface.

    The EVOLVE randomized controlled trial (EVOLVE: A Prospective Randomized Multicenter Single-blind Non-inferiority Trial to Assess the Safety and Performance of the Evolution Everolimus-Eluting Monorail Coronary Stent System [Evolution Stent System] for the Treatment of a De novo Atherosclerotic Lesion; NCT01135225) found SYNERGY to be noninferior to the durable polymer PROMUS Element everolimus-eluting stent (EES) for the angiographic end point of in-stent late lumen loss at 6 months,7 but lacked sufficient power to provide meaningful comparison(s) of clinical events.8 EVOLVE II represents the pivotal, randomized controlled clinical trial evaluating the clinical efficacy and safety of the SYNERGY stent for regulatory approval in a broad population of patients undergoing percutaneous coronary intervention (PCI).

    Methods

    EVOLVE II is a prospective, international, multicenter, randomized (1:1 SYNERGY versus PROMUS Element Plus), controlled, single-blind, noninferiority trial (EVOLVE RCT) conducted at 125 clinical sites. EVOLVE II also includes a concurrent, nonrandomized, single-arm, pharmacokinetic substudy (EVOLVE II PK), as well as a consecutively enrolled, nonrandomized, single-arm, diabetes mellitus substudy (EVOLVE II Diabetes; Figure 1), both of which will be reported separately. EVOLVE II was conducted in accordance with the US Food and Drug Administration’s Guidance for Industry E6 Good Clinical Practice: Consolidated Guidance, the Declaration of Helsinki, the International Conference on Harmonisation, and all local regulations, as appropriate. Institutional Review Boards at each center approved the study protocol and all subjects provided written informed consent. The study is registered at www.clinicalTrials.gov under identifier NCT01665053.

    Figure 1.

    Figure 1. Patient flow and disposition in the EVOLVE II Trial. CTO indicates chronic total occlusion; DS, diameter stenosis; ISR, in-stent restenosis target lesion; LM, left main; PK, pharmacokinetic; RVD, reference vessel diameter; STEMI, ST-segment–elevation myocardial infarction; and SVG, saphenous vein graft target lesion.

    Device Description

    The SYNERGY stent is a thin-strut (74–81 μm), PtCr metal alloy platform with an ultrathin (4 μmol/L) bioabsorbable Poly(D,L-lactide-co-glycolide) abluminal polymer, which elutes everolimus (100 μg/cm2). SYNERGY has been compared with the durable polymer PROMUS Element EES as described previously9,10 (Table 1).

    Table 1. Specific Design Characteristics of the SYNERGY and PROMUS Element Plus Stents

    SYNERGYPROMUS Element Plus
    Platform materialPlatinum chromiumPlatinum chromium
    Stent strut thickness74–81 μm81–86 μm
    Polymer85:15 PLGAPVDF-HFP PBMA
    Polymer typeBiodegradableBiostable
    Polymer distributionAbluminalConformal
    Polymer thickness4 μm7.8 μm
    Polymer durationApproximately 4 moPermanent
    DrugEverolimusEverolimus
    Drug/polymer ratio in active layer (wt%/wt%)45/5517/83
    Loaded drug dose100 μg/cm2100 μg/cm2

    PLGA indicates Poly(D,L-lactide-co-glycolide); and PVDF, polyvinylidene fluoride.

    Study Design and Procedure

    Eligible patients were aged ≥18 years and had either symptomatic coronary artery disease with objective evidence of ischemia or silent ischemia. Stents were implanted for treatment of ≤3 (maximum) discrete target lesions in ≤2 (maximum) 2 major epicardial vessels with lesion length ≤34 mm and reference vessel diameter ≥2.25 to ≤4.0 mm. Target stenoses were ≥50% and <100% with thrombolysis in myocardial infarction flow >1. Subjects were required to have either target stenosis ≥70% or a stenosis ≥50% to <70% with abnormal fractional flow reserve, elevated cardiac biomarkers, or objective evidence of myocardial ischemia (abnormal stress or imaging stress test). Patients with recent ST-segment–elevation myocardial infarction, left main disease, chronic total occlusions, vein graft disease, or in-stent restenosis were excluded per-protocol. Subjects who satisfied study selection criteria were randomly assigned 1:1 (stratified by diabetic status and enrollment site) to receive treatment with either SYNERGY or PROMUS Element Plus stents. Random permuted blocks were used to ensure approximate balance of treatment allocation within each stratum. EVOLVE II RCT is a single blind trial; subjects were blinded to treatment assigned and treatment received and will remain blinded until after trial completion. Packaging of the investigational control and test devices was different, therefore, the investigator performing the procedure was not blinded to the assigned treatment arm or resulting treatment. Site personnel conducting clinical follow-up, core laboratory personnel and the Clinical Events Committee were blinded to patient treatment assignment during the trial.

    DAPT with aspirin and a P2Y12 inhibitor was prescribed after PCI for at least 6 months (12 months in patients not at high risk of bleeding). An independent core laboratory evaluated all baseline and repeat angiograms (Beth Israel Deaconess Medical Center, Boston, MA). Clinical follow-up was required in-hospital at 30 days, 6, 12, and 18 months after PCI then annually between 2 and 5 years. There was no protocol-specified coronary angiography in follow-up.

    End Points

    The primary end point for EVOLVE II was the rate of 12-month target lesion failure (TLF), defined as the composite occurrence of any ischemia-driven revascularization of the target lesion, myocardial infarction (MI) related to the target vessel, or any cardiac death. Secondary clinical end points included individual components of TLF; target vessel failure defined as the composite occurrence of ischemia-driven target vessel revascularization, MI related to the target vessel or cardiac death related to the target vessel), all-cause death, and stent thrombosis (defined by the Academic Research Consortium; ST).11 All major adverse events were adjudicated by a Clinical Events Committee and the decisions of the Clinical Events Committee superseded those of the investigational center in the event of a disparity. Spontaneous MI was defined as the rise and fall of cardiac biomarkers with ≥1 value >99th percentile of the upper reference limit with evidence of myocardial ischemia. The diagnosis of periprocedural MI required at least 1 of the following: (1) CK-MB >3× upper reference limit without clinical or imaging correlates, (2) new pathological Q waves, or (3) autopsy evidence of acute MI. MI was also independently assessed by Academic Research Consortium criteria as a secondary analysis.11 Further details pertaining to the definition of MI are found in the Data Supplement document. Technical success was defined as successful delivery and deployment of the study stent to the target vessel, without balloon rupture or stent embolization with postprocedure diameter stenosis of <30% and thrombolysis in myocardial infarction 3 flow in the target lesion (as visually assessed by the treating physician). Clinical procedural success was defined as postprocedure diameter stenosis <30%, thrombolysis in myocardial infarction 3 flow in all target lesions and the absence of in-hospital MI, TVR, or cardiac death. All procedural and follow-up (through 12 months) angiograms were systematically evaluated for longitudinal stent deformation by the independent angiographic core laboratory.

    Statistical Methods

    The study primary end point, powered for noninferiority, was the rate of 12-month TLF. A 2-group Farrington–Manning test was used to test the 1-sided hypothesis of noninferiority in proportions. Specifically, if the P value from a 1-sided Farrington–Manning test was <0.025 in both the intention-to-treat and per-protocol patient populations, SYNERGY would be concluded to be noninferior to PROMUS Element Plus. This corresponds to the 1-sided upper 97.5% confidence bound for the difference in 12-month TLF rates (SYNERGY–PROMUS Element Plus) being less than the noninferiority margin. On the basis of an assumed event rate in the test (SYNERGY) and control (PROMUS Element Plus) groups of 8.0% and a noninferiority margin of 4.4%; 1684 randomized subjects (842 per group) were required (assuming a 5% attrition) to provide power (1-β) of 0.89. Continuous variables were estimated as mean±SD and compared with the Student t test. Discrete variables were reported as counts and percentages, and differences were assessed by means of the χ2 or Fisher exact tests. Cumulative event rates were estimated by the Kaplan–Meier method.

    Results

    Patients and Enrollment

    Between November 2012 and August 2013, 1684 patients were enrolled and randomized at 125 sites in North America, Europe, Australia, New Zealand, Japan, and Singapore. Of these, 846 were randomized to SYNERGY and 838 to PROMUS Element Plus (Figure 1). One-year follow-up was available in 831 (98.2%) SYNERGY and 806 (96.2%) PROMUS Element Plus stent–treated patients.

    Baseline patient clinical demographics and quantitative coronary angiographic characteristics were similar between treatment groups (Tables 2 and 3). The average age was 64 years, 31% of subjects had medically treated diabetes mellitus, more than a third had unstable angina, and more than quarter had MI diagnosed before the index PCI (Table 2). More than 20% of patients in each treatment group had multilesion (≥2 lesion) PCI and ≈75% of target lesions were classified as AHA/ACC B2/C lesion complexity (Table 3) by core laboratory adjudicated quantitative coronary angiography. Procedural characteristics (Table 4) and postprocedural angiographic results (Table 5) were similar between treatment groups with the exception that site-reported technical success was more frequent among SYNERGY-treated patients (98.3% SYNERGY versus 96.9% PROMUS Element Plus; P=0.04). Two instances of longitudinal stent deformation were observed through 12 months, both of which involved PROMUS Element Plus Stents; 1 in the PROMUS Element Plus arm and 1 in the SYNERGY arm (occurring in a PROMUS Element Plus stent mistakenly used during the index procedure). Compliance with DAPT to 6 and 12 months was 97.7% and 89.7% for SYNERGY and 96.9% and 87.3% for PROMUS Element Plus with no significant differences between stent types.

    Table 2. Baseline Characteristics of the Study Population

    Variable*SYNERGYn=846 PatientsPROMUS Element Plusn=838 PatientsP Value
    Male70.6%72.7%0.34
    Age, y, ±SD63.5±10.463.9±10.50.40
    White77.4%79.2%0.37
    Smoking, ever61.7%62.8%0.63
    Current smoker21.8%22.4%0.76
    Diabetes mellitus31.1%30.8%0.89
     Treated with insulin12.3%10.9%0.36
    Hyperlipidemia74.0%74.5%0.82
    Hypertension77.3%75.1%0.29
    Previous PCI35.8%37.3%0.52
    Previous CABG4.6%6.1%0.18
    History of CHF8.3%9.0%0.63
    Unstable angina33.9%34.8%0.69
    MI25.9%29.2%0.12

    CABG indicates coronary artery bypass graft; MI, myocardial infarction; and PCI, percutaneous coronary intervention.

    *Number (percent) based on an intent-to-treat analysis.

    P values are 2-sided and from Student t test for continuous variables and the χ2.

    Medically treated.

    Table 3. Baseline Lesion Characteristics

    SYNERGYn=1059 Lesionsn=846 PatientsPROMUS Element Plus n=1043 Lesionsn=838 PatientsP Value*
    Target lesions1.25±0.501.24±0.490.77
     2 lesions treated18.6%19.3%0.69
     3 lesions treated3.3%2.4%0.26
     >3 lesions treated0.0%0.1%0.50
    Target lesion location
     LAD41.3%41.5%0.91
     LCx25.0%26.4%0.48
     RCA33.7%32.0%0.41
     LM0.0%0.1%0.50§
    RVD, mm2.62±0.492.63±0.500.63
     RVD <2.25 mm23.9%23.3%0.76
    MLD, mm0.89±0.350.89±0.360.99
    Diameter stenosis, %66.02±12.0366.26±11.750.65
    Lesion length, mm14.09±7.5013.67±7.000.18
     Length >20 mm19.2%16.7%0.14
    Modified AHA/ACC B2/C76.8%74.3%0.19

    ACC indicates American College of Cardiology; AHA, American Heart of Association; LM, left main; MLD, minimum lumen diameter; and RVD, reference vessel diameter.

    *P values are 2-sided and from Student t test for continuous variables and the χ2 unless indicated otherwise.

    Per patient number (percent) based on an intent-to-treat analysis.

    Per lesion.

    §Fisher exact test.

    Table 4. Procedural Characteristics

    SYNERGYn=1059 Lesionsn=846 Patientsn=1011 StentsPROMUS Element Plus n=1043 Lesionsn=838 Patientsn=1079 StentsP Value*
    Technical success98.3%96.9%  0.04
    Clinical procedural success94.9%94.3%  0.56
    Stents per patient1.31±0.601.29±0.56  0.46
    Stents per target lesion1.05±0.251.04±0.25  0.32
    Total stent length implanted, mm21.45±9.0420.81±9.16  0.11
    Predilatation, %97.1%98.0%  0.18
    Postdilatation, %60.7%61.0%  0.90
    Max pressure overall, atm15.98±3.0616.09±3.13  0.41
    Longitudinal stent deformation§0.1%0.1%>0.99

    *P values are 2-sided and from Student t test for continuous variables and the χ2.

    Per lesion.

    Per patient number (percent) based on an intent-to-treat analysis.

    §Per stent.

    Occurred in a PROMUS Element Plus stent used in a SYNERGY assigned patient.

    Table 5. Postprocedural Angiographic Characteristics

    Per LesionSYNERGYn=1059 LesionsPROMUS Element Plus n=1043 LesionsP Value*
    MLD, in-stent, mm2.44±0.442.46±0.440.23
    MLD, in-segment, mm2.10±0.472.10±0.470.78
    %DS, in-stent, %7.19±9.166.55±9.710.12
    %DS, in-segment, %20.60±8.4120.93±9.130.39
    Acute gain, in-stent, mm1.55±0.451.57±0.450.33
    Acute gain, in-segment, mm1.22±0.481.21±0.470.72

    DS indicates diameter stenosis; and MLD, minimum lumen diameter.

    *P values are 2-sided and from Student’s t test for continuous variables and the χ2.

    Primary End Point Analyses

    The trial primary end point of TLF analyzed by intention-to-treat was observed in 6.7% of SYNERGY and 6.5% of PROMUS Element Plus–treated patients (1-sided 97.5% Farrington–Manning upper confidence bound of 2.68%) P=0.0005 noninferiority (Figure 2). Per-protocol analysis demonstrated TLF to be 6.4% in each treatment group (upper confidence bound, 2.51%; P=0.0003 for noninferiority). Because both intention-to-treat and per-protocol analyses demonstrate P<0.025, the SYNERGY stent is determined to be noninferior to the PROMUS Element Plus stent for TLF at 1 year. Kaplan–Meier curves for TLF event rate occurrence over time were similar for both the stents (Figure 3), as were the individual components of TLF (Figure 4). Revascularization event rates to 1 year were similar between stent platforms as well (Table 6). There were 3 definite/probable STs in the SYNERGY arm (0.4%) and 5 in the PROMUS Element Plus arm (0.6%; P=0.50). The 2 definite STs in the SYNERGY arm were acute (≤24 hours) and the 1 probable ST was subacute (6 days) postprocedure. One of the SYNERGY acute ST events involved a patient who was not treated with preprocedural aspirin. All the 5 definite/probable STs in the PROMUS Element Plus arm occurred subacutely (between 2 and 30 days; Figure 5).

    Table 6. Clinical End Points Through 12 Months

    Variable*SYNERGYn=846 PatientsPROMUS Element Plus n=838 PatientsP Value
    Death, %1.1% (9/832)1.1% (9/808)0.95
    Cardiac death, %0.5% (4/832)0.9% (7/808)0.34
    MI, %5.4% (45/832)5.0% (40/808)0.68
     Q-wave MI0.2% (2/832)0.2% (2/808)>0.99
     Non–Q-wave MI5.2% (43/832)4.7% (38/808)0.66
    TVR, %3.8% (32/832)3.6% (29/808)0.78
     PCI, %3.0% (25/832)3.2% (26/808)0.80
     CABG, %0.8% (7/832)0.4% (3/808)0.34
    TLR, %2.6% (22/832)1.7% (14/808)0.21
     PCI, %2.0% (17/832)1.7% (14/808)0.64
     CABG, %0.6% (5/832)0.0% (0/808)0.06
    Non-TLR TVR, %1.8% (15/832)2.2% (18/808)0.54
     PCI, %1.4% (12/832)1.9% (15/808)0.51
     CABG, %0.4% (3/832)0.4% (3/808)>0.99
    Stent thrombosis, %§
     Definite/probable0.4% (3/832)0.6% (5/808)0.50
     Definite0.2% (2/832)0.2% (2/808)>0.99
     Probable0.1% (1/832)0.4% (3/808)0.37

    CABG indicates coronary artery bypass graft; MI, myocardial infarction; PCI, percutaneous coronary intervention; and TLR, revascularization of the target lesion.

    *Binary event rates. Number (percent) based on an intent-to-treat analysis.

    P value from χ2 test unless otherwise noted.

    P value from Fisher exact test.

    §Stent thrombosis adjudicated according to Academic Research Consortium definition.

    Figure 2.

    Figure 2. Primary end point of target lesion failure (TLF) at 1 year. One-year TLF in the SYNERGY (blue) and PROMUS Element Plus (red) cohorts in the intent-to-treat (A) and per-protocol (B) patient populations are shown on the left. On the right, the plot shows the difference in TLF between SYNERGY and PROMUS Element Plus (black circle) with the 1-sided 97.5% Farrington–Manning upper confidence bound (UCB*) indicated by the error bar. The P values for noninferiority testing are 1-sided.

    Figure 3.

    Figure 3. Time-to-event curve for the composite primary end point of target lesion failure (TLF) through 1 year. The event rates presented here were calculated by Kaplan–Meier methodology and compared with the log-rank test. Event rate±1.5 SE. HR indicates hazard ratio (95% confidence intervals).

    Figure 4.

    Figure 4. Time-to-event curves for the components of target lesion failure through 1 year. Target lesion failure (primary end point) is a composite of any cardiac death (A), target vessel-related myocardial infarction (MI; B), and clinically indicated revascularization of the target lesion (TLR; C). Event rate±1.5 SE; P value from a log-rank test.

    Figure 5.

    Figure 5. Stent thrombosis through 12 months. Definite/probable stent thrombosis rates through 12 months. Timing is separated into acute (≤1 day; blue), early (2–30 days; green), and late (>30 days–1 year; orange). Binary event rate; P value from a χ2 test.

    Discussion

    In this pivotal trial designed to support regulatory approval of the first bioabsorbable polymer DES available in the United States, the SYNERGY stent proved to be noninferior to the PROMUS Element Plus stent for TLF at 1 year. Furthermore, rates of target vessel MI, clinically indicated/ischemia-driven revascularization of the target lesion and stent thrombosis to 1 year were low and similar for both stents.

    These clinical observations complement the finding of noninferiority of SYNERGY (versus PROMUS Element) for quantitative coronary angiographic late lumen loss (0.10 versus 0.15 mm, respectively) at 6 months reported from the EVOLVE randomized first human use trial, and provide adequate sample size from which to make more definitive conclusions about important clinical outcomes.7,8 In this regard, EVOLVE II supports the premise that the safety and efficacy of SYNERGY are at least comparable to the predicate PtCr durable polymer EES.

    Although polymer provides a reservoir for programmed drug release, it has no function after drug release is complete and may affect the late/very late safety and efficacy of DES. First generation DES polymers (EVA-BMA [SurModics, Minneapolis, MN] and SIBS-translute [Boston Scientific]) were, at times, associated with inflammation, foreign body giant cell reaction, negative vessel remodeling, and late (acquired) stent malapposition with thrombus formation.2,12,13 Durable polymers may also contribute to delayed/incomplete endothelial coverage and impaired stent healing.2,14 Although newer durable polymers may have enhanced biocompatibility and seem to be associated with improved clinical outcomes, they have still been incriminated in the occurrence of inflammation, neoatherosclerosis, and thrombosis.15,16 Indeed, neoatherosclerosis occurs earlier and with increased prevalence after DES compared with bare metal stents and has been observed with similar frequency among both first as well as newer generation DES.17 Early randomized controlled clinical trials as well as meta-analyses suggested that biodegradable polymer DES were associated with lower rates of late/very late stent thrombosis when compared with either first generation DES or bare metal stents.4,18 Conversely, more recent network meta-analyses and observational studies have suggested that the newer generation cobalt chromium (CoCr) and PtCr durable polymer (polyvinylidene fluoride) EES are associated with even lower rates of stent thrombosis when compared with other durable polymer DES, early biodegradable polymer DES, and even bare metal stents.3,1921 Furthermore, a randomized controlled clinical trial comparing the CoCr EES with its corresponding bare metal stent platform demonstrated lower rates of stent thrombosis at 1 and 2 years after primary PCI for ST-elevation MI in the CoCr EES group.21 These observations are consistent with bench and preclinical data, which suggest that the durable polyvinylidene fluoride polymer may be thromboprotective against stent thrombosis.22 Finally, a large-scale randomized controlled trial comparison of the CoCr EES versus the Nobori biodegradable polymer DES demonstrated similar long-term outcomes for both the stents.23 These apparent inconsistencies may, at least in part, be explained by differences in biodegradable polymer DES platform design.

    Both the time course and extent of endothelial stent coverage, as well as the function and maturation of endothelial cells may be influenced by multiple factors, including metal alloy, stent strut thickness, polymer composition, and distribution as well as the time course for polymer bioresorption.5,24,25 In this regard, the SYNERGY stent was designed to enhance/expedite stent healing in hopes of improving clinical outcomes by incorporating thin (74 μm) PtCr struts with an ultrathin (4 μm) Poly(D,L-lactide-co-glycolide) everolimus-eluting polymer applied only to the abluminal stent surface and which is resorbed within 4 months.7,26 The bare metal PtCr platform which remains after polymer resorption may be less proinflammatory in cell assay when compared with gold, CoCr, or cobalt nickel alloy platforms, and seems to both expedite endothelial cell stent coverage and reduce platelet adhesion when compared with PtCr covered by polyvinylidene fluoride durable polymer.25 Whether these putative preclinical attributes of the SYNERGY stent may translate into clinical benefit (reduction in stent related ischemic events or the relative treatment benefit recently observed for longer-term DAPT therapy6) will require further study.

    Limitations

    Several potential limitations to this study deserve mention. First, because the study design was single-blind (patient), physician operators were not blinded with respect to stent type deployed. Second, EVOLVE II is not adequately powered to evaluate the individual components of TLF. Third, specific complex patient and target lesion subsets were excluded from the study (ST-segment–elevation myocardial infarction, left main or saphenous vein graft target stenosis, chronic total occlusion, in-stent restenosis). Despite the lack of all-comers inclusion criteria the study population seems to reflect current clinical PCI practice with regard to the number of lesions/vessels undergoing PCI, target lesion length and angiographic complexity.27 Finally, current follow-up duration is limited to only 1 year. Indeed, longer follow-up in more complex patient/lesion subsets may better differentiate between stent platforms with different structural design or polymer-healing attributes. Previous studies comparing DES with bioresorbable and durable polymers have, at times, demonstrated progressive differences in clinical outcomes >1 year.18

    Conclusions

    The EVOLVE II randomized controlled trial demonstrates the SYNERGY coronary stent to be noninferior to the predicate PROMUS Element Plus stent for the occurrence of TLF at 1 year. Secondary end points, including ischemia-driven– revascularization of the target lesion and ST, were also similar between stents. EVOLVE II establishes comparable clinical safety and efficacy of everolimus elution from a bioresorbable polymer—PtCr metal platform when compared with the established PtCr durable polymer EES. The longer-term relative efficacy and safety of SYNERGY will be evaluated in 5-year follow-up.

    Acknowledgments

    We thank Kristine Roy, PhD (Boston Scientific Corporation), for assistance in article preparation and Songtao Jiang, MSc (Boston Scientific Corporation), for assistance with statistical analysis.

    Footnotes

    The Data Supplement is available at http://circinterventions.ahajournals.org/lookup/suppl/doi:10.1161/CIRCINTERVENTIONS.114.002372/-/DC1.

    Correspondence to Dean J. Kereiakes, MD, The Christ Hospital Heart and Vascular Center/The Lindner Research Center, 2123 Auburn Ave, Suite 424 Cincinnati, OH 45219. E-mail

    References

    • 1. Levine GN, Bates ER, Blankenship JC, Bailey SR, Bittl JA, Cercek B, Chambers CE, Ellis SG, Guyton RA, Hollenberg SM, Khot UN, Lange RA, Mauri L, Mehran R, Moussa ID, Mukherjee D, Nallamothu BK, Ting HH; American College of Cardiology Foundation; American Heart Association Task Force on Practice Guidelines; Society for Cardiovascular Angiography and Interventions. 2011 ACCF/AHA/SCAI Guideline for Percutaneous Coronary Intervention. A report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Society for Cardiovascular Angiography and Interventions.J Am Coll Cardiol. 2011; 58:e44–e122. doi: 10.1016/j.jacc.2011.08.007.CrossrefMedlineGoogle Scholar
    • 2. Virmani R, Guagliumi G, Farb A, Musumeci G, Grieco N, Motta T, Mihalcsik L, Tespili M, Valsecchi O, Kolodgie FD.Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious?Circulation. 2004; 109:701–705. doi: 10.1161/01.CIR.0000116202.41966.D4.LinkGoogle Scholar
    • 3. Kang SH, Park KW, Kang DY, Lim WH, Park KT, Han JK, Kang HJ, Koo BK, Oh BH, Park YB, Kandzari DE, Cohen DJ, Hwang SS, Kim HS.Biodegradable-polymer drug-eluting stents vs. bare metal stents vs. durable-polymer drug-eluting stents: a systematic review and Bayesian approach network meta-analysis.Eur Heart J. 2014; 35:1147–1158. doi: 10.1093/eurheartj/eht570.CrossrefMedlineGoogle Scholar
    • 4. Navarese EP, Kubica J, Castriota F, Gibson CM, De Luca G, Buffon A, Bolognese L, Margheri M, Andreotti F, Di Mario C, De Servi S.Safety and efficacy of biodegradable vs. durable polymer drug-eluting stents: evidence from a meta-analysis of randomised trials.EuroIntervention. 2011; 7:985–994. doi: 10.4244/EIJV7I8A155.CrossrefMedlineGoogle Scholar
    • 5. Palmaz JC, Bailey S, Marton D, Sprague E.Influence of stent design and material composition on procedure outcome.J Vasc Surg. 2002; 36:1031–1039.CrossrefMedlineGoogle Scholar
    • 6. Mauri L, Kereiakes DJ, Yeh RW, Driscoll-Shempp P, Cutlip DE, Steg PG, Normand SL, Braunwald E, Wiviott SD, Cohen DJ, Holmes DR, Krucoff MW, Hermiller J, Dauerman HL, Simon DI, Kandzari DE, Garratt KN, Lee DP, Pow TK, Lee PV, Rinaldi MJ, Massaro JM.Twelve or 30 months of dual antiplatelet therapy after drug-eluting stents.New Engl J Med. 2014; 371:2155–2166. doi: 10.1056/NEJMoa1409312.CrossrefMedlineGoogle Scholar
    • 7. Meredith IT, Verheye S, Dubois CL, Dens J, Fajadet J, Carrié D, Walsh S, Oldroyd KG, Varenne O, El-Jack S, Moreno R, Joshi AA, Allocco DJ, Dawkins KD.Primary endpoint results of the EVOLVE trial: a randomized evaluation of a novel bioabsorbable polymer-coated, everolimus-eluting coronary stent.J Am Coll Cardiol. 2012; 59:1362–1370. doi: 10.1016/j.jacc.2011.12.016.CrossrefMedlineGoogle Scholar
    • 8. Meredith IT, Verheye S, Weissman NJ, Barragan P, Scott D, Valdés Chávarri M, West NE, Kelbæk H, Whitbourn R, Walters DL, Kubica J, Thuesen L, Masotti M, Banning A, Sjögren I, Stables RH, Allocco DJ, Dawkins KD.Six-month IVUS and two-year clinical outcomes in the EVOLVE FHU trial: a randomised evaluation of a novel bioabsorbable polymer-coated, everolimus-eluting stent.EuroIntervention. 2013; 9:308–315. doi: 10.4244/EIJV9I3A52.CrossrefMedlineGoogle Scholar
    • 9. Menown IB, Noad R, Garcia EJ, Meredith I.The platinum chromium element stent platform: from alloy, to design, to clinical practice.Adv Ther. 2010; 27:129–141. doi: 10.1007/s12325-010-0022-9.CrossrefMedlineGoogle Scholar
    • 10. Stone GW, Teirstein PS, Meredith IT, Farah B, Dubois CL, Feldman RL, Dens J, Hagiwara N, Allocco DJ, Dawkins KD; PLATINUM Trial Investigators. A prospective, randomized evaluation of a novel everolimus-eluting coronary stent: the PLATINUM (a Prospective, Randomized, Multicenter Trial to Assess an Everolimus-Eluting Coronary Stent System [PROMUS Element] for the Treatment of Up to Two de Novo Coronary Artery Lesions) trial.J Am Coll Cardiol. 2011; 57:1700–1708. doi: 10.1016/j.jacc.2011.02.016.CrossrefMedlineGoogle Scholar
    • 11. Cutlip DE, Windecker S, Mehran R, Boam A, Cohen DJ, van Es GA, Steg PG, Morel MA, Mauri L, Vranckx P, McFadden E, Lansky A, Hamon M, Krucoff MW, Serruys PW; Academic Research Consortium. Clinical end points in coronary stent trials: a case for standardized definitions.Circulation. 2007; 115:2344–2351. doi: 10.1161/CIRCULATIONAHA.106.685313.LinkGoogle Scholar
    • 12. Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, Gold HK, Virmani R.Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization.Circulation. 2007; 115:2435–2441. doi: 10.1161/CIRCULATIONAHA.107.693739.LinkGoogle Scholar
    • 13. Vorpahl M, Finn AV, Nakano M, Virmani R.The bioabsorption process: tissue and cellular mechanisms and outcomes.EuroIntervention. 2009; 5(suppl F):F28–F35. doi: 10.4244/EIJV5IFA5.CrossrefMedlineGoogle Scholar
    • 14. Joner M, Finn AV, Farb A, Mont EK, Kolodgie FD, Ladich E, Kutys R, Skorija K, Gold HK, Virmani R.Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk.J Am Coll Cardiol. 2006; 48:193–202. doi: 10.1016/j.jacc.2006.03.042.CrossrefMedlineGoogle Scholar
    • 15. Nakazawa G, Nakano M, Otsuka F, Wilcox JN, Melder R, Pruitt S, Kolodgie FD, Virmani R.Evaluation of polymer-based comparator drug-eluting stents using a rabbit model of iliac artery atherosclerosis.Circ Cardiovasc Interv. 2011; 4:38–46. doi: 10.1161/CIRCINTERVENTIONS.110.957654.LinkGoogle Scholar
    • 16. Byrne RA, Joner M, Kastrati A.Polymer coatings and delayed arterial healing following drug-eluting stent implantation.Minerva Cardioangiol. 2009; 57:567–584.MedlineGoogle Scholar
    • 17. Otsuka F, Pacheco E, Perkins LE, Lane JP, Wang Q, Kamberi M, Frie M, Wang J, Sakakura K, Yahagi K, Ladich E, Rapoza RJ, Kolodgie FD, Virmani R.Long-term safety of an everolimus-eluting bioresorbable vascular scaffold and the cobalt-chromium XIENCE V stent in a porcine coronary artery model.Circ Cardiovasc Interv. 2014; 7:330–342. doi: 10.1161/CIRCINTERVENTIONS.113.000990.LinkGoogle Scholar
    • 18. Serruys PW, Farooq V, Kalesan B, de Vries T, Buszman P, Linke A, Ischinger T, Klauss V, Eberli F, Wijns W, Morice MC, Di Mario C, Corti R, Antoni D, Sohn HY, Eerdmans P, Rademaker-Havinga T, van Es GA, Meier B, Jüni P, Windecker S.Improved safety and reduction in stent thrombosis associated with biodegradable polymer-based biolimus-eluting stents versus durable polymer-based sirolimus-eluting stents in patients with coronary artery disease: final 5-year report of the LEADERS (Limus Eluted From A Durable Versus ERodable Stent Coating) randomized, noninferiority trial. JACC Cardiovasc Interv. 2013; 6:777–789. doi: 10.1016/j.jcin.2013.04.011.CrossrefMedlineGoogle Scholar
    • 19. Navarese EP, Tandjung K, Claessen B, Andreotti F, Kowalewski M, Kandzari DE, Kereiakes DJ, Waksman R, Mauri L, Meredith IT, Finn AV, Kim HS, Kubica J, Suryapranata H, Aprami TM, Di Pasquale G, von Birgelen C, Kedhi E.Safety and efficacy outcomes of first and second generation durable polymer drug eluting stents and biodegradable polymer biolimus eluting stents in clinical practice: comprehensive network meta-analysis.BMJ. 2013; 347:f6530.CrossrefMedlineGoogle Scholar
    • 20. Palmerini T, Biondi-Zoccai G, Della Riva D, Stettler C, Sangiorgi D, D’Ascenzo F, Kimura T, Briguori C, Sabatè M, Kim HS, De Waha A, Kedhi E, Smits PC, Kaiser C, Sardella G, Marullo A, Kirtane AJ, Leon MB, Stone GW.Stent thrombosis with drug-eluting and bare-metal stents: evidence from a comprehensive network meta-analysis.Lancet. 2012; 379:1393–1402. doi: 10.1016/S0140-6736(12)60324-9.CrossrefMedlineGoogle Scholar
    • 21. Sabaté M, Räber L, Heg D, Brugaletta S, Kelbaek H, Cequier A, Ostojic M, Iñiguez A, Tüller D, Serra A, Baumbach A, von Birgelen C, Hernandez-Antolin R, Roffi M, Mainar V, Valgimigli M, Serruys PW, Jüni P, Windecker S.Comparison of newer-generation drug-eluting with bare-metal stents in patients with acute ST-segment elevation myocardial infarction: a pooled analysis of the EXAMINATION (clinical Evaluation of the Xience-V stent in Acute Myocardial INfArcTION) and COMFORTABLE-AMI (Comparison of Biolimus Eluted From an Erodible Stent Coating With Bare Metal Stents in Acute ST-Elevation Myocardial Infarction) trials.JACC Cardiovasc Interv. 2014; 7:55–63. doi: 10.1016/j.jcin.2013.07.012.CrossrefMedlineGoogle Scholar
    • 22. Kolandaivelu K, Swaminathan R, Gibson WJ, Kolachalama VB, Nguyen-Ehrenreich KL, Giddings VL, Coleman L, Wong GK, Edelman ER.Stent thrombogenicity early in high-risk interventional settings is driven by stent design and deployment and protected by polymer-drug coatings.Circulation. 2011; 123:1400–1409. doi: 10.1161/CIRCULATIONAHA.110.003210.LinkGoogle Scholar
    • 23. Natsuaki M, Kozuma K, Morimoto T, Kadota K, Muramatsu T, Nakagawa Y, Akasaka T, Igarashi K, Tanabe K, Morino Y, Ishikawa T, Nishikawa H, Awata M, Abe M, Okada H, Takatsu Y, Ogata N, Kimura K, Urasawa K, Tarutani Y, Shiode N, Kimura T; NEXT Investigators. Biodegradable polymer biolimus-eluting stent versus durable polymer everolimus-eluting stent: a randomized, controlled, noninferiority trial.J Am Coll Cardiol. 2013; 62:181–190. doi: 10.1016/j.jacc.2013.04.045.CrossrefMedlineGoogle Scholar
    • 24. Eppihimer M.Impact of polymer type and location on stent thrombogenicity and endothelial cell coverage. EuroIntervention. 2013; EuroPCR Abstracts 2013: Abstract 341.Google Scholar
    • 25. Eppihimer MJ, Sushkova N, Grimsby JL, Efimova N, Kai W, Larson S, Forsyth B, Huibregtse BA, Dawkins KD, Wilson GJ, Granada JF.Impact of stent surface on thrombogenicity and vascular healing: a comparative analysis of metallic and polymeric surfaces.Circ Cardiovasc Interv. 2013; 6:370–377. doi: 10.1161/CIRCINTERVENTIONS.113.000120.LinkGoogle Scholar
    • 26. Wilson GJ, Huibregtse BA, Pennington DE, Dawkins KD.Comparison of the SYNERGY with the PROMUS (XIENCE V) and bare metal and polymer-only Element control stents in porcine coronary arteries.EuroIntervention. 2012; 8:250–257. doi: 10.4244/EIJV8I2A39.CrossrefMedlineGoogle Scholar
    • 27. Yeh RW, Czarny MJ, Normand SL, Kereiakes DJ, Holmes DR, Brindis RG, Weaver WD, Rumsfeld JS, Roe MT, Kim S, Driscoll-Shempp P, Mauri L.Evaluating the generalizability of a large streamlined cardiovascular trial: comparing hospitals and patients in the dual antiplatelet therapy study versus the National Cardiovascular Data Registry.Circ Cardiovasc Qual Outcomes. 2015; 8:96–102. doi: 10.1161/CIRCOUTCOMES.114.001239.LinkGoogle Scholar