Cardioversion Efficacy Using Pulsed Biphasic or Biphasic Truncated Exponential Waveforms: A Randomized Clinical Trial
Abstract
Background
Several different defibrillators are currently used for cardioversion and defibrillation of cardiac arrhythmias. The efficacy of a novel pulsed biphasic (PB) waveform has not been compared to other biphasic waveforms. Accordingly, this study aims to compare the efficacy and safety of PB shocks with biphasic truncated exponential (BTE) shocks in patients undergoing cardioversion of atrial fibrillation or ‐flutter.
Methods and Results
This prospective, randomized study included patients admitted for elective direct current cardioversion. Patients were randomized to receive cardioversion using either PB or BTE shocks. We used escalating shocks until sinus rhythm was obtained or to a maximum of 4 shocks. Patients randomized to PB shocks received 90, 120, 150, and 200 J and patients randomized to BTE shocks received 100, 150, 200, and 250 J, as recommended by the manufacturers. In total, 69 patients (51%) received PB shocks and 65 patients (49%) BTE shocks. Successful cardioversion, defined as sinus rhythm 4 hours after cardioversion, was achieved in 43 patients (62%) using PB shocks and in 56 patients (86%) using BTE shocks; ratio 1.4 (95% CI 1.1–1.7) (P=0.002). There was no difference in safety (ie, myocardial injury judged by changes in high‐sensitive troponin I levels; ratio 1.1) (95% CI 1.0–1.3), P=0.15. The study was terminated prematurely because of an adverse event.
Conclusions
Cardioversion using a BTE waveform was more effective when compared with a PB waveform. There was no difference in safety between the 2 waveforms, as judged by changes in troponin I levels.
Clinical Trial Registration
Introduction
Atrial fibrillation (AF) is the most common cardiac arrhythmia and affects millions of people worldwide.1 To reduce symptoms and prevent disease progression, direct current cardioversion is a widely used procedure for patients with AF.2 During the past decades, biphasic waveforms have proven superior to monophasic waveforms for cardioversion of supraventricular tachycardia in several randomized studies.3, 4, 5, 6 In these studies, more patients were restored to sinus rhythm when receiving biphasic shocks compared with monophasic shocks, using less total delivered energy. Currently, several different biphasic waveforms are available in commercial defibrillators.
Recently, a new pulsed biphasic (PB) waveform was approved for direct current cardioversion and defibrillation.7 This waveform delivers a chopped modulation of a biphasic truncated exponential (BTE) waveform, where current is rapidly and repeatedly turned on and off. In a nonrandomized clinical study, this PB waveform was found to be more effective in cardioversion of AF or atrial flutter compared with a monophasic waveform.8 The PB waveform is designed to deliver a high peak and average current at low energy levels,9, 10 and it is hypothesized that the waveform can deliver high shock efficacy at low energy levels.7, 11
To date, there are no clinical data on the efficacy of the impedance‐compensated (ie, adjusted to patient's chest impedance) PB waveform, which is currently in use for direct current cardioversion and defibrillation. Lately, international guidelines have identified an important knowledge gap on using the PB waveform, and state that clinical data are warranted.12 Because the PB waveform has never been compared to other biphasic waveforms in a randomized clinical study, we accordingly aimed to compare the cardioversion efficacy and safety of a PB waveform with a BTE waveform.
Methods
Study Design and Setting
We conducted a prospective, randomized study using a 2×2 factorial design. The study randomized patients to cardioversion using either PB or BTE shocks and investigated the use of 100% oxygen versus room air during the cardioversion procedure. The study was conducted at the Regional Hospital of Randers, Denmark, in an outpatient clinic performing elective day case cardioversions of patients with AF or atrial flutter.
Intervention
Patients were randomized to receive cardioversion using either PB shocks (Multipulse Biowave®, Schiller Defigard 5000, Schiller AG, Baar, Switzerland) or BTE shocks (LIFEPAK 20, Medtronic/Physio‐Control Inc., Redmond, WA). Both waveforms use impedance compensation, where the patient's impedance is measured by a sensing pulse prior to shock delivery. In addition, the PB waveform uses the impedance to calculate an appropriate pulse‐to‐pause ratio. Diagrams of the waveforms are presented in Figure 1.

Patients were randomized using simple randomization with random numbers from 1 to 4 in sealed envelopes. Before cardioversion, the envelope was opened, assigning patients to 1 of the 4 treatment groups: (1) PB shocks and 100% oxygen, (2) PB shocks and room air, (3) BTE shocks and 100% oxygen, or (4) BTE shocks and room air.
Study Population and Ethics
All patients admitted for elective cardioversion of AF or atrial flutter were eligible for inclusion. Patients were included at an elective visit 1 to 2 days before cardioversion. The exclusion criteria were age <18 years, pregnancy, untreated hyperthyroidism, or an oxygen saturation <92% and supraventricular arrhythmias other than AF or atrial flutter. All patients were required to be adequately anticoagulated or alternatively have undergone a recent transesophageal echocardiography documenting the absence of intracardiac thrombi.13
Information on patient characteristics was obtained at the precardioversion check including the patient's height, weight, blood pressure, heart rate, oxygen saturation (on room air), and a 12‐lead ECG. Symptoms were assessed according to the European Heart Rhythm Association score of AF‐related symptoms.13 All data on medications and comorbidities were retrieved from the patient's medical records corresponding to the day before cardioversion.
Oral and written informed consent was obtained from all patients at the precardioversion visit. The study was conducted in accordance with national requirements and the principles of the Declaration of Helsinki. The study was approved by The National Committee on Health Research Ethics (no. 1‐10‐72‐150‐13) and the Danish Data Protection Agency (no. 1‐16‐02‐425‐13).
End Points
The primary end point was successful cardioversion, defined as sinus rhythm 4 hours after cardioversion. Secondary end points included sinus rhythm 1‐ and 30 minutes after cardioversion. The safety of the 2 defibrillators was evaluated by measuring high‐sensitive cardiac troponin I (Hs‐cTnI) before and 4 hours after cardioversion. Complications, in terms of cardioversion‐induced arrhythmias, were further assessed including cases of ventricular tachyarrhythmia, short‐duration asystole <20 s, and transient bradycardia <45 min−1. Data were obtained from ECG recordings and telemetric surveillance. Furthermore, any other complications including skin burns were noted in the patient's journal by the treating nurse.
Cardioversion Protocol
Cardioversion shocks were provided until sinus rhythm was restored or to a maximum of 4 shocks. The patients randomized to PB shocks received 90, 120, 150, and 200 J, and the patients randomized to BTE shocks received 100, 150, 200, and 250 J. We used these escalating shock protocols based on manufacturer's recommended settings.
Anesthesia was provided in a standardized fashion to patients using 1 mg intravenous propofol per kilogram body weight to a maximum dose of patient's height in centimeters minus 100 cm. Subsequent boluses of 20 mg were administered as required. During the procedure, patients were treated with either oxygen or room air to maintain oxygen saturations of 94% to 98%.
We used the defibrillator‐specific self‐adhesive wet polymer gel pads. All shocks were delivered unblinded using an anterior–posterior pad position.14 The anterior pad was applied to the left of the sternum with the pad centrum at third intercostal space. The posterior pad was placed with the pad edge laterally to the spinal column on the lower part of the scapula. The defibrillators were switched to “R” wave synchronized mode.
One minute after last shock delivery and after 30 minutes, the cardiac rhythm was documented by obtaining a 12‐lead ECG. After 4 hours with continuous cardiac telemetric surveillance, another 12‐lead ECG was recorded.
Blood Sampling and Biochemical Analysis
Blood samples were drawn from a venous catheter into lithium heparin tubes and centrifuged immediately at 344g for 15 minutes. High‐sensitive cardiac troponin I (hs‐cTnI) was analyzed in a DANAK ISO 15189 accredited laboratory using ARCHITECT STAT hs‐cTnI assay (Abbott Laboratories, Abbott Park, IL). The coefficient of variation was maximum 10% at the limit of quantification (10 ng/L).
Statistical Analysis
In this study, we assumed a 15% difference in efficacy (ie, the proportion of patients in sinus rhythm 4 hours after cardioversion), corresponding to an efficacy of 95% for 1 of the waveforms, and 80% for the other waveform, as previously used.15 To achieve a power of 80% to detect this difference (ie, to reject the null hypothesis of no difference between the waveforms), a sample size of 75 patients in each group was needed. Normally distributed data were expressed as mean±SD, non‐normally distributed data were expressed as median (25‐ and 75‐percentile), and categorical variables were expressed as numbers (percentages). We reported the primary and secondary end point as a ratio between treatment groups with 95% CIs. When comparing cardioversion success, we used the χ2 test or Fisher's exact test. We separately analyzed patients with AF and atrial flutter, and patients receiving and not receiving amiodarone. To evaluate the hs‐cTnI values and the assumptions behind the test, we used X‐Y plots, Q‐Q plots, and Bland–Altman plots. The Bland–Altman plots were performed separately for the 2 waveform groups using the hs‐cTnI values before and after cardioversion. The assumptions were met on logarithm transformed data. Accordingly, the hs‐cTnI values were transformed to the logarithmic scale, and differences were calculated. Subsequently, the differences on the logarithmic scale were transformed to the original scale, and consequently reported as a ratio of geometric means. The shock protocols were not completely identical, because of different available energy settings in each device, and the difference in success between the 2 waveforms was therefore further evaluated according to the energies used. Furthermore, we compared cardioversion success between patients treated with 100% oxygen and room air by binomial generalized linear models with a log link function including and excluding the additional intervention. Moreover, we tested for interaction between the 2 types of intervention. We used likelihood ratio tests for these models.
The graphical illustration was performed using Graph‐Pad Prism version 6 and for statistical computing, we used R statistical software, version 3.1.1.16
Results
Patients
A total of 134 patients were enrolled between September 2013 and August 2014 (Figure 2). Of these, 69 patients (51%) were randomized to cardioversion using PB shocks, and 65 patients (49%) to BTE shocks.

Baseline characteristics of the 134 patients included are presented in Table 1. The 2 treatment groups were balanced on patient's demographics, comorbidities, cardiac medications, vital parameters, laboratory data, and AF‐related symptoms, except that more patients in the BTE waveform group received amiodarone than patients in the PB waveform group. In total, 35 (50.7%) of patients randomized to PB shocks received 100% oxygen, and 32 (49.2%) of patients randomized to BTE shocks received 100% oxygen.
Patient's Characteristics | PB Waveform | BTE Waveform |
---|---|---|
Demographics | ||
No. of patients | 69 | 65 |
Age, y (mean±SD) | 66±9 | 67±8 |
Male (%) | 51 (74) | 51 (78) |
Atrial flutter | 9 (13) | 9 (14) |
AF or ‐flutter duration in months, median (quartiles) | 3 (2; 9) | 5 (2; 24) |
Prior cardioversions, median (range) | 0 (0–5) | 0 (0–6) |
Prior pulmonary vein ablation, median (range) | 0 (0–3) | 0 (0–1) |
Comorbidities, % | ||
Hypertension | 51 (74) | 51 (78) |
Congestive heart failure | 20 (29) | 12 (19) |
Valvular heart disease | 3 (4) | 7 (11) |
Thyroid disease | 0 (0) | 2 (3) |
Prior stroke/transient ischemic attack | 6 (9) | 4 (6) |
Prior myocardial infarction | 1 (1) | 5 (8) |
Ischemic heart disease | 13 (19) | 12 (18) |
Hyperlipidemia | 29 (42) | 28 (43) |
Chronic obstructive pulmonary disease | 2 (3) | 5 (8) |
Cardiac medication, % | ||
Amiodarone | 6 (9) | 18 (28) |
Digoxin | 14 (19) | 18 (28) |
Flecainide | 1 (1) | 1 (2) |
ACE/angiotensin‐II blocker | 44 (64) | 40 (62) |
β‐Blocker | 57 (83) | 53 (82) |
Calcium antagonist | 16 (23) | 20 (31) |
Vital parameters and laboratory data | ||
Body mass index, kg/m2 | 29±6 | 30±6 |
Systolic blood pressure, mm Hg | 136±17 | 135±21 |
Diastolic blood pressure, mm Hg | 82±14 | 80±13 |
Heart rate before cardioversion, min−1 | 80±20 | 80±18 |
Estimated glomerular filtration rate | 69±14 | 69±17 |
Hemoglobin, mmol/L | 9.3±0.8 | 9.3±0.9 |
Thyroid‐stimulating hormone (IU ×10−3/L) | 1.57 (1.00; 2.59) | 1.67 (0.99; 2.49) |
Total propofol dose, mg | 113±43 | 107±33 |
Patient's symptoms (EHRA score), % | ||
I | 31 (45) | 26 (40) |
II | 31 (45) | 29 (45) |
III | 7 (10) | 10 (15) |
IV | 0 (0) | 0 (0) |
ACE indicates angiotensin‐converting enzyme; AF, atrial fibrillation; BTE, biphasic truncated exponential; EHRA, European Heart Rhythm Association score of AF‐related symptoms; PB, pulsed biphasic.
Shock Characteristics
In total, 192 PB shocks and 139 BTE shocks were delivered. The median (25‐ and 75‐percentile) number of shocks delivered per patient was for the PB waveform 3 (1 and 4) and for the BTE waveform 2 (1 and 3). The corresponding median total energy delivered was for PB shocks 360 J (90 and 560 J) and for BTE shocks 250 J (100 and 450 J). The mean (±SD) transthoracic impedance at first shock was for PB shocks 73±14 Ω and for BTE shocks 72±15 Ω.
End Points
The primary end point and secondary efficacy end points are presented in Table 2. The subgroup analyses of the primary end point are presented in Table 3 (patient's characteristics of the subgroups can be found in Tables S1 and S2).
Successful Cardioversions | PB Waveform | BTE Waveform | Ratio (95% CI) | P Value |
---|---|---|---|---|
Primary end point | ||||
Sinus rhythm after 4 hours | 43 (62) | 56 (86) | 1.4 (1.1–1.7) | 0.002 |
Secondary end points | ||||
Sinus rhythm after 1 minute | 46 (67) | 58 (89) | 1.3 (1.1–1.6) | 0.002 |
Sinus rhythm after 30 minutes | 46 (67) | 57 (88) | 1.3 (1.1–1.6) | 0.004 |
Data are numbers of successful cardioversions (%) and estimates are reported with their corresponding 95% CI. BTE indicates biphasic truncated exponential; PB, pulsed biphasic.
PB Waveform | BTE Waveform | Ratio (95% CI) | P Value | |
---|---|---|---|---|
Sinus rhythm after 4 hours | ||||
Atrial fibrillation patients (n=116) | 35 (58) | 47 (84) | 1.4 (1.1–1.8) | 0.002 |
Atrial flutter patients (n=18) | 7 (77) | 9 (100) | 1.3 (0.9–1.9) | 0.13 |
Patient receiving amiodarone (n=24) | 2 (33) | 16 (89) | 2.7 (0.9–8.4) | 0.05 |
Patients not receiving amiodarone (n=110) | 40 (64) | 40 (85) | 1.4 (1.1–1.7) | 0.01 |
Data are numbers of successful cardioversions (%) and estimates are reported with their corresponding 95% CI. BTE indicates biphasic truncated exponential; PB, pulsed biphasic.
The cumulative success rate at the different energy levels used is shown in Figure 3. As shocks were delivered at 150 and 200 J for both waveforms, we compared cardioversion success at these energies. For 150‐J shocks, the PB waveform success rate was 26% and 40% for the BTE waveform (ratio; 1.5, 95% CI 0.8–2.9, P=0.2). For shocks at 200 J, the PB shock success rate was 30% and 44% for the BTE waveform (ratio: 1.5, 95% CI 0.7–3.0, P=0.3). There was no statistical difference in demographic data among patients in whom cardioversion was attempted at 150 and 200 J (data not shown). Moreover, we tested for interaction between the cardioversion efficacy of the 2 waveforms and the 100% oxygen/room air intervention. The interaction between the interventions was not significant (P=0.3) and, furthermore, there was no difference between 100% oxygen and room air on cardioversion efficacy either with (P=0.46) or without (P=0.84) adjustment for the waveform used.

Defibrillator Safety
There was no difference in hs‐cTnI change between PB and BTE shocks; ratio 1.1 (95% CI, 1.0–1.3, P=0.15). The overall risk of any complication following a cardioversion shock was for the PB waveform 7 cases out of 192 shocks (4%) and for BTE waveform 4 cases out of 139 shocks (3%); ratio 1.3 (95% CI, 0.4–4.2, P=0.77). Two patients (3%) treated with PB shocks developed ventricular tachyarrhythmia immediately after cardioversion compared to none when using BTE shocks (see below). Periods of short‐duration asystole (all <20 s) were reported for 1 patient in the PB group (1%) and 2 patients (3%) in the BTE group. Transient bradycardia (<45 min−1) was observed in 4 patients (6%) from the PB group and 2 patients (3%) from the BTE group. There were no observations of skin burns following cardioversion using either the pulsed or truncated exponential waveform during the study.
Adverse Events
Because of a major adverse event, the study was terminated prematurely. The adverse event occurred in a 61‐year‐old male AF patient who was randomized to cardioversion using PB shocks. The patient had a history of hypertension and hypercholesterolemia, but without any history of structural heart disease or heart failure. The patient had a normal QT‐interval, plasma potassium, and creatinine levels within normal range, and did not receive any antiarrhythmic medication. A synchronized cardioversion shock of 90 J was delivered, and immediately after the patient developed sustained polymorphic ventricular tachycardia (Figure 4). Three stacked shocks with the PB device did not result in successful defibrillation. After 8 minutes of advanced life support, including further repeated defibrillation attempts, the patient was switched to the BTE defibrillator and successfully defibrillated at first shock (300 J). After return of spontaneous circulation, the patient was transferred to an invasive cardiac center for acute angiography, showing no signs of coronary artery disease. The patient was treated with therapeutic hypothermia for 24 hours. Analysis of the rhythm strip printed by the defibrillator revealed that the intended synchronized cardioversion shock was delivered asynchronously (Figure 4).

Three months before this event, we experienced a case where a patient, after shock delivery, developed a short run of nonsustained ventricular tachycardia, which was rapidly converted with an additional shock. The rhythm strip revealed that a nonsynchronous shock had been administered. The physician who performed the cardioversion felt confident that the PB device was set to synchronized mode; however, the delivery of the asynchronous shock was at that time interpreted as caused by a human error.
Following these incidents, we became aware that the synchronization failure may have been attributable to an error in the PB device rather than human errors. Consequently, after the second case emerged, we tested the defibrillator's synchronization function in our animal laboratory, confirming that asynchronous shocks were delivered despite that the defibrillator was switched into synchronous mode.
The PB device was returned to the manufacturer for further analysis, confirming that an error in the device software caused the delivery of the asynchronous shock despite that the PB device was set in synchronous modus. All required reporting to national regulatory agencies and the manufacturer of the PB device was performed and subsequently a safety notice was published on the synchronization software failure.17
Discussion
This is the first randomized clinical study comparing the efficacy of a PB waveform with a BTE waveform demonstrating a higher efficacy for BTE shocks compared with PB shocks in cardioverting AF. There was no difference in safety (ie, myocardial injury measured by changes in troponin I levels).
For every fourth patient cardioverted with BTE shocks, 1 patient would have failed cardioversion if the PB waveform had been used. This substantial amount of failed cardioversions when using PB shocks may have profound clinical implications. The “failed” patients suffered longer time in arrhythmia, may have to be readmitted for another cardioversion attempt, possibly prescribed a concomitant antiarrhythmic drug, or treated with rate control therapy.
The impedance‐compensated (ie, adjusted to differences in chest impedance) PB waveform, which is in clinical use, has not previously been compared to other waveforms. In cardioversion of atrial arrhythmias, a nonimpedance compensated PB waveform was more efficient than a monophasic waveform on the energy used to obtain sinus rhythm in patients with atrial arrhythmias.8 One study on 104 out‐of‐hospital cardiac arrest patients with shockable rhythm compared a nonimpedance compensated version of the PB waveform to a BTE waveform.18 The study reported similar defibrillation success rates for PB shocks (90.4%) and the BTE waveform (weighted average 91.8%).
Different versions of PB waveforms have been evaluated in animal studies using ventricular fibrillation pig models.19, 20 The studies found that the same energy levels were needed to terminate ventricular fibrillation using PB shocks compared to a rectilinear biphasic waveform,19 and more energy when compared to a BTE waveform.20 These findings are consistent with our study, where more energy was required to cardiovert when using PB shocks compared with BTE shocks.
Importantly, the success rate of BTE shocks in our study (89% 1‐minute postcardioversion and 86% at discharge) was in accordance with other studies. Accordingly, a study comparing BTE shocks to rectilinear biphasic shocks report a success rate of 91%,15 and similarly a success rate of 90% was found in a study comparing BTE shocks with monophasic shocks; both studies used a comparable shock energy protocol and end point (1‐minute postcardioversion).6 Furthermore, previous studies have compared other biphasic waveforms in cardioversion of AF and atrial flutter. A rectilinear waveform was compared to a BTE waveform, but no studies have reported a difference between these waveforms in cardioversion success.15, 21, 22, 23
In this study, PB shocks were delivered up to the maximal energy output at 200 J allowed by the device. For BTE shocks, the maximum energy output used in this study was 250 J rather than the possible maximum 360 J. This was in order to not disadvantage the PB waveform and to allow a reasonable comparison between the 2 waveforms by maintaining parity in escalating energies between groups. Delivery of the BTE shocks at a maximal energy of 360 J may further increase the cardioversion success rate.
In this study, there was no difference in myocardial injury between waveforms and no changes in hs‐cTnI were observed regardless of waveform and energy used. These findings are in accordance with previous studies, showing no increase in troponin I following cardioversion.21, 24, 25, 26, 27, 28 In our study, the risk of postshock arrhythmias was low, although we did experience 2 cases of ventricular arrhythmia. The use of a low‐energy escalating protocol is currently recommended for cardioversion,2, 13 to minimize, eg, postcardioversion arrhythmias. However, a retrospective study reviewing 6398 cardioversion procedures found 5 cases of ventricular fibrillation, in which the energy selection was 50 to 100 J.29 The risk of inducing ventricular arrhythmia after cardioversion may be increased when using low‐energy shocks close to the defibrillation threshold, which correspond to the upper limit of vulnerability for inducing ventricular fibrillation.30 Furthermore, a study found a higher first shock success and no increase in postshock arrhythmias when using a fixed high‐energy biphasic shock protocol (200 J) compared with a low‐energy escalating shock protocol (100–150–200 J).31 Importantly, using higher energy selection for cardioversion may be considered to decrease the risk of induced ventricular arrhythmia, although the optimal shock energy protocol for biphasic cardioversion remains to be identified.
Limitations
Ideally, the PB and BTE waveforms should be tested within the same device using the same defibrillation electrodes and exactly identical shock protocols. Importantly, the results of this study reflect the clinical use of the equipment, using the available energy settings in each device, allowing a reasonable comparison of the 2 waveforms. The study was terminated earlier than expected, with 16 patients below the calculated sample size. However, the actual difference in cardioversion success was much higher than expected according to the sample size estimation. This was because of a software failure where asynchronous shocks caused 2 cases of ventricular arrhythmia. It is well known that asynchronous shocks can cause this harm if applied in the vulnerable phase of the T‐wave. However, it is unknown if shock delivery at other positions in the heart cycle might affect cardioversion efficacy. Lastly, despite randomization, treatment with amiodarone was unevenly distributed in the 2 groups, but the amiodarone‐stratified subgroup analysis did not change the overall result of the study.
Conclusions
This study demonstrated that when compared by energy levels, cardioversion using a BTE waveform was more effective compared to a PB waveform. There was no difference in myocardial injury between the 2 waveforms (ie, changes in troponin I levels).
Sources of Funding
This study was funded by the Regional Hospital of Randers, Denmark and Marie de Lancy Pedersen's Foundation, Denmark. The PB defibrillator used in this study was provided free of charge from Schiller AG, Baar, Switzerland including the appropriate defibrillation pads. Physio‐Control Inc., Redmond, WA also provided defibrillation pads free of charge from the Danish distributor.
Disclosures
Deakin served as the immediate past chair of ILCOR Advanced Life Support taskforce. The remaining authors have no disclosures to report.
Acknowledgments
The authors acknowledge Mariann Frost Rasmussen, RN; The Heart Clinic, Department of Internal Medicine, Regional Hospital of Randers; Department of Anesthetics, Regional Hospital of Randers; and Clinical Research Unit, Regional Hospital of Randers, Denmark.
Supplemental Material
Table S1. Characteristics of the 116 Patients With Atrial Fibrillation and the 18 Patients With Atrial Flutter
Table S2. Characteristics of the 24 Patients Receiving Amiodarone Treatment and the 110 Patients Not Receiving Amiodarone
- Download
- 178.77 KB
References
1.
Chugh SS, Havmoeller R, Narayanan K, Singh D, Rienstra M, Benjamin EJ, Gillum RF, Kim YH, McAnulty JH Jr, Zheng ZJ, Forouzanfar MH, Naghavi M, Mensah GA, Ezzati M, Murray CJ. Worldwide epidemiology of atrial fibrillation: a Global Burden of Disease 2010 Study. Circulation. 2014;129:837–847.
2.
January CT, Wann LS, Alpert JS, Calkins H, Cleveland JC Jr, Cigarroa JE, Conti JB, Ellinor PT, Ezekowitz MD, Field ME, Murray KT, Sacco RL, Stevenson WG, Tchou PJ, Tracy CM, Yancy CW. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2014;64:e1–e76.
3.
Koster RW, Dorian P, Chapman FW, Schmitt PW, O'Grady SG, Walker RG. A randomized trial comparing monophasic and biphasic waveform shocks for external cardioversion of atrial fibrillation. Am Heart J. 2004;147:e20.
4.
Mittal S, Ayati S, Stein KM, Schwartzman D, Cavlovich D, Tchou PJ, Markowitz SM, Slotwiner DJ, Scheiner MA, Lerman BB. Transthoracic cardioversion of atrial fibrillation: comparison of rectilinear biphasic versus damped sine wave monophasic shocks. Circulation. 2000;101:1282–1287.
5.
Ambler JJ, Deakin CD. A randomized controlled trial of efficacy and ST change following use of the Welch‐Allyn MRL PIC biphasic waveform versus damped sine monophasic waveform for external DC cardioversion. Resuscitation. 2006;71:146–151.
6.
Page RL, Kerber RE, Russell JK, Trouton T, Waktare J, Gallik D, Olgin JE, Ricard P, Dalzell GW, Reddy R, Lazzara R, Lee K, Carlson M, Halperin B, Bardy GH; BiCard Investigators . Biphasic versus monophasic shock waveform for conversion of atrial fibrillation: the results of an international randomized, double‐blind multicenter trial. J Am Coll Cardiol. 2002;39:1956–1963.
7.
Cansell A. Impulses or a series of impulses for defibrillation and device to generate them. US Patent. 2000; 6.493.580.
8.
Krasteva V, Trendafilova E, Cansell A, Daskalov I. Assessment of balanced biphasic defibrillation waveforms in transthoracic atrial cardioversion. J Med Eng Technol. 2001;25:68–73.
9.
Kostov J, Mudrov T, Dotsinsky I, Mudrov N. Comparison between two defibrillation waveforms. J Med Eng Technol. 2010;34:429–436.
10.
Krasteva V, Cansell A, Daskalov IK. Transthoracic defibrillation with chopping‐modulated biphasic waveforms. J Med Eng Technol. 2001;25:163–168.
11.
White RD. Waveforms for defibrillation and cardioversion: recent experimental and clinical studies. Curr Opin Crit Care. 2004;10:202–207.
12.
Callaway CW, Soar J, Aibiki M, Bottiger BW, Brooks SC, Deakin CD, Donnino MW, Drajer S, Kloeck W, Morley PT, Morrison LJ, Neumar RW, Nicholson TC, Nolan JP, Okada K, O'Neil BJ, Paiva EF, Parr MJ, Wang TL, Witt J; Advanced Life Support Chapter Collaborators . Part 4: advanced life support: 2015 international consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Circulation. 2015;132:S84–S145.
13.
European Heart Rhythm Association, European Association for Cardio‐Thoracic Surgery, Camm AJ, Kirchhof P, Lip GY, Schotten U, Savelieva I, Ernst S, Van Gelder IC, Al‐Attar N, Hindricks G, Prendergast B, Heidbuchel H, Alfieri O, Angelini A, Atar D, Colonna P, De Caterina R, De Sutter J, Goette A, Gorenek B, Heldal M, Hohloser SH, Kolh P, Le Heuzey JY, Ponikowski P, Rutten FH. Guidelines for the management of atrial fibrillation: the Task Force for the Management of Atrial Fibrillation of the European Society of Cardiology (ESC). Eur Heart J. 2010;31:2369–2429.
14.
Kirchhof P, Eckardt L, Loh P, Weber K, Fischer RJ, Seidl KH, Bocker D, Breithardt G, Haverkamp W, Borggrefe M. Anterior‐posterior versus anterior‐lateral electrode positions for external cardioversion of atrial fibrillation: a randomised trial. Lancet. 2002;360:1275–1279.
15.
Deakin CD, Connelly S, Wharton R, Yuen HM. A comparison of rectilinear and truncated exponential biphasic waveforms in elective cardioversion of atrial fibrillation: a prospective randomized controlled trial. Resuscitation. 2013;84:286–291.
16.
R Core Team . R: A Language and Environment for Statistical Computing. Vienna, Austria: R Foundation for Statistical Computing; 2014. Available at: http://www.R-project.org/. Accessed October 1, 2014.
17.
Schiller Medical, Safety Notice Defigard 5000. Health Products Regulation Authority. 2014. Available at: http://www.hpra.ie/docs/default-source/field-safety-notices/October-2014/v22248_fsn.pdf?sfvrsn=2. Accessed November 10, 2016.
18.
Didon JP, Fontaine G, White RD, Jekova I, Schmid JJ, Cansell A. Clinical experience with a low‐energy pulsed biphasic waveform in out‐of‐hospital cardiac arrest. Resuscitation. 2008;76:350–353.
19.
Li Y, Wang H, Cho JH, Didon JP, Bisera J, Weil MH, Tang W. Comparison of efficacy of pulsed biphasic waveform and rectilinear biphasic waveform in a short ventricular fibrillation pig model. Resuscitation. 2009;80:1047–1051.
20.
Sullivan JL, Melnick SB, Chapman FW, Walcott GP. Porcine defibrillation thresholds with chopped biphasic truncated exponential waveforms. Resuscitation. 2007;74:325–331.
21.
Neal S, Ngarmukos T, Lessard D, Rosenthal L. Comparison of the efficacy and safety of two biphasic defibrillator waveforms for the conversion of atrial fibrillation to sinus rhythm. Am J Cardiol. 2003;92:810–814.
22.
Kim ML, Kim SG, Park DS, Gross JN, Ferrick KJ, Palma EC, Fisher JD. Comparison of rectilinear biphasic waveform energy versus truncated exponential biphasic waveform energy for transthoracic cardioversion of atrial fibrillation. Am J Cardiol. 2004;94:1438–1440.
23.
Alatawi F, Gurevitz O, White RD, Ammash NM, Malouf JF, Bruce CJ, Moon BS, Rosales AG, Hodge D, Hammill SC, Gersh BJ, Friedman PA. Prospective, randomized comparison of two biphasic waveforms for the efficacy and safety of transthoracic biphasic cardioversion of atrial fibrillation. Heart Rhythm. 2005;2:382–387.
24.
Piechota W, Gielerak G, Ryczek R, Kazmierczak A, Bejm J, Piechota W. Cardiac troponin I after external electrical cardioversion for atrial fibrillation as a marker of myocardial injury—a preliminary report. Kardiol Pol. 2007;65:664–669; discussion 670‐1.
25.
Allan JJ, Feld RD, Russell AA, Ladenson JH, Rogers MA, Kerber RE, Jaffe AS. Cardiac troponin I levels are normal or minimally elevated after transthoracic cardioversion. J Am Coll Cardiol. 1997;30:1052–1056.
26.
Lund M, French JK, Johnson RN, Williams BF, White HD. Serum troponins T and I after elective cardioversion. Eur Heart J. 2000;21:245–253.
27.
Bonnefoy E, Chevalier P, Kirkorian G, Guidolet J, Marchand A, Touboul P. Cardiac troponin I does not increase after cardioversion. Chest. 1997;111:15–18.
28.
Vikenes K, Omvik P, Farstad M, Nordrehaug JE. Cardiac biochemical markers after cardioversion of atrial fibrillation or atrial flutter. Am Heart J. 2000;140:690–696.
29.
Gallagher MM, Yap YG, Padula M, Ward DE, Rowland E, Camm AJ. Arrhythmic complications of electrical cardioversion: relationship to shock energy. Int J Cardiol. 2008;123:307–312.
30.
Hwang C, Swerdlow CD, Kass RM, Gang ES, Mandel WJ, Peter CT, Chen PS. Upper limit of vulnerability reliably predicts the defibrillation threshold in humans. Circulation. 1994;90:2308–2314.
31.
Glover BM, Walsh SJ, McCann CJ, Moore MJ, Manoharan G, Dalzell GW, McAllister A, McClements B, McEneaney DJ, Trouton TG, Mathew TP, Adgey AA. Biphasic energy selection for transthoracic cardioversion of atrial fibrillation. The BEST AF Trial. Heart. 2008;94:884–887.
Information & Authors
Information
Published In
Copyright
© 2017 The Authors. Published on behalf of the American Heart Association, Inc., by Wiley Blackwell. This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial‐NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.
History
Received: 14 October 2016
Accepted: 9 February 2017
Published in print: March 2017
Published online: 8 March 2017
Keywords
Subjects
Notes
(J Am Heart Assoc. 2017;6:e004853. https://doi.org/10.1161/JAHA.116.004853.)
Authors
Funding Information
Regional Hospital of Randers
Marie de Lancy Pedersen's Foundation
Metrics & Citations
Metrics
Citations
Download Citations
If you have the appropriate software installed, you can download article citation data to the citation manager of your choice. Select your manager software from the list below and click Download.
- What Mistakes Can Be Made When Performing the Electrical Cardioversion Procedure?—Analysis of Emergency Medical Team Performance during the Championships in Emergency Medicine, Healthcare, 12, 17, (1724), (2024).https://doi.org/10.3390/healthcare12171724
- Impact of Defibrillator Electrode Placement on Outcome of Electrical Cardioversion of Atrial Fibrillation: A Pilot Observational Study, Journal of the American Heart Association, 13, 13, (2024)./doi/10.1161/JAHA.123.034817
- Association Between Defibrillation Using LIFEPAK 15 or ZOLL X Series and Survival Outcomes in Out‐of‐Hospital Cardiac Arrest: A Nationwide Cohort Study, Journal of the American Heart Association, 13, 7, (2024)./doi/10.1161/JAHA.123.033913
- First human safety and effectiveness study of defibrillation with a novel patch wearable cardioverter-defibrillator, Europace, 26, 7, (2024).https://doi.org/10.1093/europace/euae189
- External electrical and pharmacological cardioversion for atrial fibrillation, atrial flutter or atrial tachycardias: a network meta-analysis, Cochrane Database of Systematic Reviews, 2024, 6, (2024).https://doi.org/10.1002/14651858.CD013255.pub2
- Defibrillation Strategies for Refractory Ventricular Fibrillation, New England Journal of Medicine, 388, 9, (860-863), (2023).https://doi.org/10.1056/NEJMc2216382
- Fokus Notfallmedizin 2022/2023 – Zusammenfassung ausgewählter notfallmedizinischer StudienFocus emergency medicine 2022/2023—Summary of selected studies in emergency medicine, Die Anaesthesiologie, 72, 11, (809-820), (2023).https://doi.org/10.1007/s00101-023-01330-z
- A Randomized Comparison of Delivered Energy in Cardioversion of Atrial Fibrillation: Biphasic Truncated Exponential Versus Pulsed Biphasic Waveforms, Diagnostics, 11, 6, (1107), (2021).https://doi.org/10.3390/diagnostics11061107
- Anterior–Lateral Versus Anterior–Posterior Electrode Position for Cardioverting Atrial Fibrillation, Circulation, 144, 25, (1995-2003), (2021)./doi/10.1161/CIRCULATIONAHA.121.056301
- Balanced Energy/Time Optimal Defibrillation, ASME Letters in Dynamic Systems and Control, 2, 2, (2021).https://doi.org/10.1115/1.4052271
- See more
Loading...
View Options
Login options
Check if you have access through your login credentials or your institution to get full access on this article.
Personal login Institutional LoginPurchase Options
Purchase this article to access the full text.
eLetters(0)
eLetters should relate to an article recently published in the journal and are not a forum for providing unpublished data. Comments are reviewed for appropriate use of tone and language. Comments are not peer-reviewed. Acceptable comments are posted to the journal website only. Comments are not published in an issue and are not indexed in PubMed. Comments should be no longer than 500 words and will only be posted online. References are limited to 10. Authors of the article cited in the comment will be invited to reply, as appropriate.
Comments and feedback on AHA/ASA Scientific Statements and Guidelines should be directed to the AHA/ASA Manuscript Oversight Committee via its Correspondence page.