Can Transcranial Ultrasonication Increase Recanalization Flow With Tissue Plasminogen Activator?

Originally publishedhttps://doi.org/10.1161/01.STR.0000013789.15436.42Stroke. 2002;33:1399–1404

    Abstract

    Background and Purpose In thrombolytic therapy for acute ischemic stroke, it is essential to obtain rapid thrombolysis before ischemic neuronal injury occurs. To develop a new technique of thrombolysis for acute ischemic stroke, the effect of transcranially applied ultrasound (TUS) on thrombolysis was examined.

    Methods An occlusion model of rabbit femoral artery was produced with thrombin after establishment of stenotic flow and endothelial damage. After stable occlusion was confirmed, monteplase (mtPA) was administered intravenously, and ultrasound (490 kHz, 0.13 W/cm2) was applied through a piece of temporal bone (TUS group; n=9). The control group received mtPA alone (tissue plasminogen activator [tPA] group; n=12). To verify the efficacy of TUS, femoral artery flow was measured during the procedure.

    Results The recanalization ratio was 16.7% (2 of 12) in the tPA group and 66.7% (6 of 9) in the TUS group. The recanalization ratio in the TUS group was higher than that in the tPA group (P=0.03). Patency flow ratio, which was defined as recanalization flow divided by baseline flow, of the TUS group (44.6±13.9%) was significantly greater than that of the tPA group (9.9±6.8%) at 60 minutes (P=0.025). Patency flow ratio became higher in the TUS group than in the tPA group between 20 and 30 minutes from the start of thrombolysis.

    Conclusions Low-frequency and low-intensity TUS enhanced thrombolysis by mtPA in a rabbit femoral artery occlusion model. This technique should be clinically useful for thrombolysis in acute ischemic stroke.

    Current thrombolytic therapy for acute ischemic stroke in general involves the intravenous administration of thrombolytic agents1 or intra-arterial local fibrinolysis with the technique of radiological intervention.2 Most prospective clinical trials have been performed in acute ischemic stroke within 3 hours3,4 because delayed use of thrombolytic agents and delayed recanalization of major vessels may increase the risk of intracerebral hemorrhage. Moreover, recanalization rates by intravenous tissue plasminogen activator (tPA) range from 10% for internal carotid artery occlusion to 30% for middle cerebral artery occlusion.5 In the Prolyse in Acute Cerebral Thromboembolism (PROACT II) study, the recanalization rate was 66% for the recombinant prourokinase group.6 It has become an essential condition of the thrombolytic treatment of acute ischemic stroke that thrombolysis should be completed before ischemic neuronal injury occurs; therefore, it is important to dissolve the thrombus as early as possible.

    In relation to thrombolytic therapy of myocardial infarction, there have been reports on the thrombolytic effects of ultrasound (US) both in vitro and in vivo (Table 1).7–21 US thrombolytic therapy with the use of an intravascular catheter has also been performed, which leads to earlier recanalization.22,23 We have previously reported that low-frequency, low-intensity US penetrates the cranium and enhances thrombolysis in vitro.7 This fact led us to begin to develop a new technique of thrombolytic therapy for acute ischemic stroke that would make possible thrombolysis of a cerebral embolism at an extremely early stage. Thrombolysis with transcranially applied US (TUS) has been investigated only in in vitro experiments. This is the first in vivo study of thrombolysis with low-frequency, low-intensity TUS.

    Characteristic of Published Reports of Thrombolysis With US

    AuthorProtocolFrequencyOutputModeThrombolytic AgentThrombolysis Ratio*Reflow Ratio or % Flow Ratio*Sonication Time, min
    CW indicates continuous wave; PW, pulsed wave; and interval, intermittent insonation.
    *Compared with control results.
    Francis19 (1992)In vitro1 MHz1–8 W/cm2CWtPA140% (1 W/cm2)60
    150% (2 W/cm2)60
    178% (4 W/cm2)60
    458% (8 W/cm2)60
    Lauer13 (1992)In vitro1 MHz1.75 W/cm2IntervaltPA152%200
    Luo14 (1993)In vitro1 MHz1–2.2 W/cm2CWStreptokinase55% (2.2 W/cm2)30
    Harpaz10 (1994)In vitro1 MHz2.5 W/cm2CWtPA514.50%40
    Kimura11 (1994)In vitro300 kHz, 1 MHz0.07, 0.4 W/cm2CWtPA
    Francis9 (1995)In vitro1 MHz4 W/cm2PWtPA147% (tPA uptake)60
    193% (tPA uptake)240
    Akiyama7 (1998)In vitro211.5 kHz, 1.03 MHz0.25 W/cm2CWUrokinase140%240
    Suchkova21 (1998)40 kHz0.25–1.5 W/cm2CWtPA175% (0.25 W/cm2)60
    290% (0.75 W/cm2)60
    375% (1 W/cm2)60
    385% (1.5 W/cm2)60
    Behrens8 (1999)In vitro33.3, 71.4 kHz0.5–3.4 W/cm2tPA130% (33 kHz)60
    152% (33 kHz)180
    143% (71 kHz)60
    171% (71 kHz)180
    Kudo20 (1989)In vivo200 kHztPA400%
    Lauer13 (1992)In vivo1 MHz1.75 W/cm2IntervaltPA916%180
    Kornowski12 (1994)In vivo1 MHz6.3 W/cm2IntervaltPA120
    Luo15 (1996)In vivo26 kHz16 W/cm2Streptokinase983%30
    Riggs17 (1997)In vivo1 MHz2 W/cm2IntervalStreptokinase410%120
    Luo16 (1998)In vivo37 kHz160 WPWStreptokinase1492%60
    Suchkova18 (2000)In vivo40 kHz0.75 W/cm2CWStreptokinase1185%120

    Materials and Methods

    Animal Model

    All animal procedures were performed under the guidance of the Animal Research Committee (Jikei University School of Medicine, Tokyo, Japan). Twenty-one New Zealand White rabbits weighing 2400 to 3200 g were used. The rabbits were anesthetized with intravenous thiopental sodium (20 mg/kg) via the marginal ear vein. A tracheal intubation was performed, and the anesthesia was maintained with 1% to 1.5% isoflurane. Arterial blood pressure was monitored during the procedure. The animals were kept normothermic (38.0±0.3°C) by use of a heating blanket, and core body temperature with rectal digital thermometer was measured.

    Thrombotic occlusion of the femoral artery was produced by an approach that has been described previously.12,17 An ultrasonic blood flowmeter (Transonic 1RB; Transonic) was set in place proximal to the femoral artery, and the peripheral end of a saphenous artery was ligated. Prestenotic flow was recorded via ultrasonic flowmeter (T206, Transonic) connected to a computer running PowerLab data acquisition software (Chart, ADI Instruments), and the arterial flow was monitored during the procedure. Stenosis was produced distal to the flow probe by constricting the artery with 5-0 silk suture to reduce the flow, and baseline flow was measured. Then a 1-cm segment of the femoral artery distal to the stenosis was clamped with temporary aneurysm clips, and the segment was traumatized by 3 external compressions with blunt forceps to produce endothelial injury. Then a polyethylene (PE50) catheter (Intramedic, Nippon Becton Dickinson Co Ltd) filled with 100 U/mL bovine thrombin (T-4648, Sigma Chemical Co) was inserted from the medial saphenous artery to just distal to the stenosis. To produce clots, 15 to 20 μL of blood was suctioned, and the blood was reintroduced with 1 U thrombin. Immediately after the PE50 catheter was withdrawn, the femoral artery distal to the stenosis was clamped. The proximal and distal clips were released 30 minutes later, and the absence of blood flow was monitored with the flow probe for 10 minutes to confirm a stable occlusion of the femoral artery. If the blood flow was not 0 mL/min at this time, additional clamping was performed for 10 minutes. All animals were observed for 10 minutes after the clamps were removed. The total age of the clots was 40 or 50 minutes. Each experiment was commenced after confirmation of zero flow.

    TUS System and Temperature Measurement

    The US system was described in our previous study.7 The automatic intermittent controller was installed as a new addition to the previous system. The US transducer used in this study was a ceramic type with a diameter of 10 mm, newly designed for animal studies. The transducer was connected to an amplifier, and an US generator (Honda Denshi Co) provided a continuous-wave output at 490.6 kHz. A human temporal bone (4 mm thick) was placed just below the transducer and wrapped in transparent kitchen plastic sheeting (Saran Wrap, Asahi Chemical) and used in a model example of TUS (Figure 1). US was applied with a simple cooling device to prevent heating of the temporal bone by an US transducer, which could affect the blood vessels. This device was made of a plastic container with transparent plastic sheeting adhering to its floor so as not to attenuate the US emissions. The container was filled with saline at room temperature, and the wrapped bone and the US transducer were inserted into it. The skin was not closed over the femoral artery. The incised edge of the operating field was drawn up with a silk thread, and the field was then filled with saline at room temperature. The TUS apparatus was placed in the water tank so that the end of its plastic container was 10 mm above the site of the embolism. The saline in the pool was exchanged when suitable to avoid temperature rise around the arteries. At the time of insonation, the temperature within the muscle surrounding the femoral artery was monitored continuously with a needle-type temperature sensor (PTW-100A, Unique Medical) and was taken as an indication of a rise in the temperature of the acoustic field of the transducer. We have already reported that low-frequency US has good penetration through the temporal bone.7 On the basis of that finding, we selected a frequency of 490.6 kHz for thrombolysis with TUS. The US intensity was determined with an Ultrasound Power Meter (UPM-DT-1; Ohmic Instruments Co). When the US intensity was set at 0.83 W/cm2, the intensity through the temporal bone was 0.13 W/cm2. We did not measure the spatial power distribution of US beam by hydrophone because the total intensity 0.83 W/cm2, which equals approximately 0.011 MPa in water, is less than the threshold level of cavitation at 490.6 kHz.24,25 To avoid tissue heating, US was applied in cycles of 2 minutes with a 30-second pause for 10 minutes. Then US was not applied for 5 minutes. These 10-minute cycles, followed by 5-minute breaks, were repeated 4 times. Total intermittent application time of US was, with pauses, 60 minutes. Actual application time of US was 32 minutes. The blood flow and the intramuscular temperature were measured during the 30-second pauses, and these measurements were repeated every 2 minutes.

    Figure 1. Scheme of experiment site. The saline was physiological saline at room temperature. The distance from the ultrasound transducer to the thrombus was 50 mm. A needle-type transducer for temperature measurement was inserted into the muscle surrounding the femoral artery within the acoustic field of the transducer.

    Study Protocol

    After embolization was completed, monteplase (mtPA) (Cleactor, Eisai Co) was administered into the posterior auricular vein. Two groups were used: the TUS group received 1.2 mg mtPA with TUS, and the tPA group received 1.2 mg mtPA alone. TUS was performed simultaneously with the administration of mtPA, and the blood flow was measured continuously for 60 minutes. The blood flow and intramuscular temperature were recorded every 2 minutes. In this study, patency flow ratio was defined as recanalization flow divided by baseline flow. Recanalization was defined as an increase of at least 0.5 mL/min in the blood flowmeter reading, whether or not blood flow 60 minutes after the procedure was patent. Blood flow was measured every 2 minutes, and the aggregate over 60 minutes was taken as the flow volume. At the end of the experiment, cardiac arrest was induced in all of the animals by administration of an overdose of pentobarbital, and samples for histological examination were excised and fixed by 10% buffered formalin. The specimens were stained with hematoxylin and eosin.

    Statistical Analysis

    All statistical analyses were performed with the use of SAS software. All data were expressed as mean±SEM. Fisher’s exact test was used to compare the occurrence of recanalization between the 2 groups (Table 2). Statistical analysis of the patency flow ratio and basic parameters was performed by tailed t test (Table 3, Figure 2). The time course of patency flow ratio was analyzed by repeated-measures ANOVA (Figure 3). A value of P<0.05 was considered significant.

    Basic Parameters in TUS Group and tPA Group

    tPATUS
    Values are mean±SEM. Basic parameters showed no significant difference between the 2 groups.
    n129
    Body weight, g2858±452766±29
    Baseline flow, mL/min7.0±1.05.3±0.6
    Stenosis ratio, %20.1±2.323.7±1.3

    Occurrence of Recanalization in TUS Group and tPA Group

    GroupnRecanalization Ratio, %P
    Any recanalization (complete or partial flow improvement) was found at 60 minutes in 2 of 12 arteries in the tPA group and 6 of 9 in the TUS group. Recanalization ratio in the TUS groups was significantly higher than that in the tPA group (P=0.03).
    tPA1216.7
    TUS966.70.03

    Figure 2. Effect of TUS on thrombolysis with the use of tPA in a rabbit femoral artery occlusion model. Patency flow ratio at 60 minutes is shown, expressed as percentage of baseline flow. Patency flow ratio in the TUS group was significantly higher than that in the control group (TPA) (P=0.025, tailed t test). Values are mean±SEM.

    Figure 3. Time course of thrombolytic effect of TUS. Patency flow ratio during the observation period in the TUS group and control group (TPA) is shown, expressed as percentage of baseline flow. Between 52 and 60 minutes, patency flow ratio in the TUS group was significantly higher than that in the tPA group (P=0.03, repeated-measures ANOVA). Each data point represents mean±SEM of number of rabbits in each group.

    Results

    Embolization of the femoral artery was performed in 21 rabbits with a mean body weight of 2809±25 g and without significant intergroup differences in weight. Their mean blood pressure during experiments was 66.5±2.3 mm Hg, and no significant intergroup difference was found in this variable. Mean rectal temperatures was 38.0±0.3°C. The mean baseline flow was 8.1±0.9 mL/min. The mean stenosis ratio in the tPA group was 20.1±2.3% and in the TUS group was 23.7±1.3% (Table 2), which was not significantly different between the 2 groups. The recanalization ratio was 16.7% (2 of 12) in the tPA group and 66.7% (6 of 9) in the TUS group. The recanalization ratio in the TUS group was thus higher than that in the tPA group (P=0.03) (Table 3). The patency flow ratio in the TUS group (44.6±13.9%) was significantly greater than that in the tPA group (9.9±6.8%) (P=0.025) (Figure 2). The patency flow ratio became significantly higher with TUS between 52 and 60 minutes from the start of thrombolysis (Figure 3). Arteries that showed complete recanalization after TUS were found to be patent with small focal residual thrombus in light microscopic examination (Figure 4A [patent vessel] and 4B [partially patent vessel]). There were no additional endothelial injuries or histological changes in arteries to suggest damage by US exposure. Increase in intramuscular temperature was <3°C in the TUS group (Figure 5).

    Figure 4. Histological examination of rabbit femoral arteries after TUS. A, In a rabbit femoral artery that was patent after 60 minutes of TUS, residual thrombus is seen in the lumen without evidence of vessel wall damage. B, In a rabbit femoral artery that was partially patent after the procedure, residual thrombus obstructs the vessel (hematoxylin and eosin stain; magnification ×20 for A and B; bar=250 μm for A and B).

    Figure 5. Time course of thermal effect of TUS in the muscle surrounding the femoral artery. Temperature during the observation period in the TUS group and control group (TPA) is shown. In the TUS group, the temperature increase was <3°C. Each data point represents mean±SEM.

    Discussion

    The present study clearly demonstrated that low-frequency (490.6 kHz) and low-intensity (0.13 W/cm2) US, which was applied through the human temporal bone, significantly enhanced thrombolysis by mtPA in a rabbit femoral artery occlusion model. This in vivo study on thrombolysis in TUS extends previously performed in vitro studies.7,8,18 Various thrombolytic therapies, including intravenous administration of tPA1–5 and intra-arterial application of prourokinase,6 have been reported for acute ischemic stroke. Delayed thrombolysis causes irreversible neuronal injury and carries risk of hemorrhagic complications.2,5 This study sought to develop a prompt and less invasive therapy for acute ischemic stroke.

    We modified the femoral artery occlusion model,12,17 which consisted of stenosis with endothelial injury and autologous thrombus. Justification of animal models of acute ischemic stroke should be evaluated on the basis of stenosis, endothelial injury, and thrombus formation. Stenosis of arteries is the fundamental requirement of a cerebral embolism model. The stenosis in this study (mean value of 24.3% stenosis by flow volume) was suitable for an experimental occlusion model based on the hemodynamic theory.26,27 Endothelial injury is also essential to thrombus formation, and therefore we injured the endothelium to lodge the autologous thrombus, which was injected externally in the femoral artery.

    Various frequencies have been used for thrombolysis with US; the majority of studies report the use of 1 MHz (Table 1). Our selection of low-frequency (490 kHz) US was based on the following considerations. First, low-frequency US can penetrate the skull well and has the ability to accelerate thrombolysis by urokinase.7 Because tissue penetration declines at higher-frequency US, lower-frequency US is ideal for use in TUS thrombolytic therapy. Second, we sought to avoid adverse effects on the brain tissue. The mechanical effect of US is proportional to peak rarefactional pressure but is inversely proportional to US frequency.24 In addition, it is generally believed that the mechanical biological effect has a threshold according to which no effect occurs unless a certain output level is exceeded.25 The relation between rarefactional pressure and US frequency is defined as p2/f=constant, where p is rarefactional pressure and f is frequency. Furthermore, the mechanical effect of US has a threshold according to which no effect occurs under a certain output level.25 On the basis of this equation, the rarefactional pressure due to low frequencies, such as 30 and 70 kHz used in a previous study,8 is approximately 3 times that which occurs at the 490.6-kHz frequency. High rarefactional pressure may cause adverse effects such as cavitation.25 Additionally, we selected the intensity (0.13 W/cm2) to be as low as possible to avoid the mechanical biological effect and to dissolve thrombus by minimal invasion.

    In addition to the mechanical biological effect, the thermal biological effect caused by the following 2 phenomena should also be considered an adverse effect. The first phenomenon is the temperature rise in the brain tissue and the skull by the US absorption phenomenon. According to the standard of the American Institute of Ultrasound in Medicine,25 the degree of thermal biological effect can be evaluated by the thermal index (TI), which indicates the temperature increase per unit volume of human tissue. In our experimental conditions of 490.6 KHz and intermittent application of US at 0.83 W/cm2, TI in the brain tissue was estimated at 1.01 for 0.13 W/cm2, and TI in the cranium was 2.08 for 0.83 W/cm2; these parameters were chosen to avoid the thermal adverse effect in the clinical application. Since the value of TI in soft tissue increases proportionally with frequency increase, the thermal effect can be decreased in lower-frequency US. Therefore, it is possible, in the interest of safety, to choose an optimal US frequency range that can avoid a large thermal effect in higher-frequency US and a large mechanical effect in lower-frequency US under the same power level. The other thermal effect is caused by the US transducer surface, which induced the energy loss in the transducer. By the use of a simple cooling system, our system succeeded in restricting the temperature rise to <3°C in the muscle surrounding the femoral artery, as shown in Figure 5. However, since this temperature increase was not enough to avoid this adverse thermal effect completely, a more precise cooling system should be used between the 2 surfaces of human temporal bone and the US probe in clinical practice.

    The mechanism underlying enhanced thrombolysis by US is not fully understood. Because US alone does not enhance thrombolysis, as reported previously,14 it is thought that the enhancing effect of US on thrombolysis is not caused by direct destruction of the thrombus but rather by accelerated transport of thrombolytic agents. Several studies have suggested that permeation and disaggregation of fibrin fibers contribute to this mechanism.28–31 It has been verified from many in vitro studies that the thrombolysis ratio rises to approximately 1.5 times that obtained without US. In contrast, most of the in vivo studies have demonstrated that the enhancing effect is markedly stronger than that observed with the in vitro studies (Table 1). Our results also showed much greater enhancement of thrombolysis: when US was applied with mtPA, the recanalization ratio increased 5 times, and the patency ratio after 60 minutes was 20 times greater than that with mtPA alone. These differences between in vitro and in vivo results may be caused not only by direct thrombolytic effect but also by other related factors, such as endothelial function, platelet activation, and altered metabolism.32

    In considering therapeutic applications for acute ischemic stroke, it is important to choose optimum US frequency, intensity, and sonication methods. In regard to tPA used in clinical practice, Alexandrov et al33 recently reported that patients in whom recanalization was monitored with transcranial Doppler had better outcome than those in whom transcranial Doppler was not used, and they reported the possibility of facilitated thrombolysis with combined tPA and US. However, 2-MHz and 0.2-W US in transcranial Doppler should be markedly attenuated to <5% through the cranium,34 and there has been no direct evidence for the enhancing effect of US at megahertz frequency on thrombolysis. One critical problem to be addressed is that of the limitation of TUS treatment in terms of safety. Further studies are needed to examine the ischemic brain to optimize US frequency and intensity as well as to develop new US equipment with a cooling system acceptable for clinical use.

    We conclude that low-frequency, low-intensity TUS enhances thrombolysis with mtPA and improves blood perfusion. This method might be promising for shortening the period of recanalization and reducing the risk of hemorrhagic complications.

    This work was supported in part by grant H11-001 from the Research on Advanced Technology program of the Ministry of Health and Welfare. The authors gratefully acknowledge Yoshihiro Sakamoto for help in performing experiments.

    Footnotes

    Correspondence to Toshihiro Ishibashi, MD (E-mail ), Department of Neurosurgery, or Hiroshi Furuhata, MD, PhD (E-mail [email protected]), Department of Medical Engineering Laboratory, Jikei University School of Medicine, 3-25-8 Nishi-Shinbashi, Minato-ku, Tokyo 105-8461, Japan.

    References

    • 1 NINDS rt-PA Stroke Study Group. Tissue plasminogen activator for acute ischemic stroke. N Engl J Med. 1995; 333: 1581–1587.CrossrefMedlineGoogle Scholar
    • 2 Ernst R, Pancioli A, Tomsick T, Kissela B, Woo D, Kanter D, Jauch E, Carrozzella J, Spilker J, Broderick J. Combined intravenous and intra-arterial recombinant tissue plasminogen activator in acute ischemic stroke. Stroke. 2000; 31: 2552–2557.CrossrefMedlineGoogle Scholar
    • 3 Albers GW, Bates VE, Clark WM, Bell R, Verro P, Hamilton SA. Intravenous tissue-type plasminogen activator for treatment of acute stroke: the Standard Treatment with Alteplase to Reverse Stroke (STARS) study. JAMA. 2000; 283: 1145–1150.CrossrefMedlineGoogle Scholar
    • 4 Katzan IL, Furlan AJ, Lloyd LE, Frank JI, Harper DL, Hinchey JA, Hammel JP, Qu A, Sila CA. Use of tissue-type plasminogen activator for acute ischemic stroke: the Cleveland area experience. JAMA. 2000; 283: 1151–1158.CrossrefMedlineGoogle Scholar
    • 5 Wolpert SM, Bruckmann H, Greenlee R, Wechsler L, Pessin MS, del Zoppo GJ, for the rt-PA Acute Stroke Study Group. Neuroradiologic evaluation of patients with acute stroke treated with recombinant tissue plasminogen activator. AJNR Am J Neuroradiol. 1993; 14: 3–13.MedlineGoogle Scholar
    • 6 Furlan A, Higashida R, Wechsler L, Gent M, Rowley H, Kase C, Pessin M, Ahuja A, Callahan F, Clark WM, Silver F, Rivera F. Intra-arterial prourokinase for acute ischemic stroke: the PROACT II study: a randomized controlled trial: Prolyse in Acute Cerebral Thromboembolism. JAMA. 1999; 282: 2003–2011.CrossrefMedlineGoogle Scholar
    • 7 Akiyama M, Ishibashi T, Yamada T, Furuhata H. Low-frequency ultrasound penetrates the cranium and enhances thrombolysis in vitro. Neurosurgery. 1998; 43: 828–832; comment 832–833.CrossrefMedlineGoogle Scholar
    • 8 Behrens S, Daffertshofer M, Spiegel D, Hennerici M. Low-frequency, low-intensity ultrasound accelerates thrombolysis through the skull. Ultrasound Med Biol. 1999; 25: 269–273.CrossrefMedlineGoogle Scholar
    • 9 Francis CW, Blinc A, Lee S, Cox C. Ultrasound accelerates transport of recombinant tissue plasminogen activator into clots. Ultrasound Med Biol. 1995; 21: 419–424.CrossrefMedlineGoogle Scholar
    • 10 Harpaz D, Chen X, Francis CW, Meltzer RS. Ultrasound accelerates urokinase-induced thrombolysis and reperfusion. Am Heart J. 1994; 127: 1211–1219.CrossrefMedlineGoogle Scholar
    • 11 Kimura M, Iijima S, Kobayashi K, Furuhata H. Evaluation of the thrombolytic effect of tissue-type plasminogen activator with ultrasonic irradiation: in vitro experiment involving assay of the fibrin degradation products from the clot. Biol Pharm Bull. 1994; 17: 126–130.CrossrefMedlineGoogle Scholar
    • 12 Kornowski R, Meltzer RS, Chernine A, Vered Z, Battler A. Does external ultrasound accelerate thrombolysis? Results from a rabbit model. Circulation. 1994; 89: 339–344.CrossrefMedlineGoogle Scholar
    • 13 Lauer CG, Burge R, Tang DB, Bass BG, Gomez ER, Alving BM. Effect of ultrasound on tissue-type plasminogen activator–induced thrombolysis. Circulation. 1992; 86: 1257–1264.CrossrefMedlineGoogle Scholar
    • 14 Luo H, Steffen W, Cercek B, Arunasalam S, Maurer G, Siegel RJ. Enhancement of thrombolysis by external ultrasound. Am Heart J. 1993; 125: 1564–1569.CrossrefMedlineGoogle Scholar
    • 15 Luo H, Nishioka T, Fishbein MC, Cercek B, Forrester JS, Kim CJ, Berglund H, Siegel RJ. Transcutaneous ultrasound augments lysis of arterial thrombi in vivo. Circulation. 1996; 94: 775–778.CrossrefMedlineGoogle Scholar
    • 16 Luo H, Birnbaum Y, Fishbein MC, Peterson TM, Nagai T, Nishioka T, Siegel RJ. Enhancement of thrombolysis in vivo without skin and soft tissue damage by transcutaneous ultrasound. Thromb Res. 1998; 89: 171–177.CrossrefMedlineGoogle Scholar
    • 17 Riggs PN, Francis CW, Bartos SR, Penney DP. Ultrasound enhancement of rabbit femoral artery thrombolysis. Cardiovasc Surg. 1997; 5: 201–207.CrossrefMedlineGoogle Scholar
    • 18 Suchkova VN, Baggs RB, Francis CW. Effect of 40-kHz ultrasound on acute thrombotic ischemia in a rabbit femoral artery thrombosis model: enhancement of thrombolysis and improvement in capillary muscle perfusion. Circulation. 2000; 101: 2296–2301.CrossrefMedlineGoogle Scholar
    • 19 Francis CW, Onundarson PT, Carstensen EL, Blinc A, Meltzer RS, Schwarz K, Marder VJ. Enhancement of fibrinolysis in vitro by ultrasound. J Clin Invest. 1992; 90: 2063–2068.CrossrefMedlineGoogle Scholar
    • 20 Kudo S. Thrombolysis with ultrasound effect [in Japanese]. Tokyo Jikeikai Med J. 1989; 104: 1005–1012.Google Scholar
    • 21 Suchkova V, Siddiqi FN, Carstensen EL, Dalecki D, Child S, Francis CW. Enhancement of fibrinolysis with 40-kHz ultrasound. Circulation. 1998; 98: 1030–1035.CrossrefMedlineGoogle Scholar
    • 22 Rosenschein U, Gaul G, Erbel R, Amann F, Velasguez D, Stoerger H, Simon R, Gomez G, Troster J, Bartorelli A, Pieper M, Kyriakides Z, Laniado S, Miller HI, Cribier A, Fajadet J. Percutaneous transluminal therapy of occluded saphenous vein grafts: can the challenge be met with ultrasound thrombolysis? Circulation. 1999; 99: 26–29.CrossrefMedlineGoogle Scholar
    • 23 Rosenschein U, Furman V, Kerner E, Fabian I, Bernheim J, Eshel Y. Ultrasound imaging–guided noninvasive ultrasound thrombolysis: preclinical results. Circulation. 2000; 102: 238–245.CrossrefMedlineGoogle Scholar
    • 24 Apfel RE, Holland CK. Gauging the likelihood of cavitation from short-pulse, low-duty cycle diagnostic ultrasound. Ultrasound Med Biol. 1991; 17: 179–185.CrossrefMedlineGoogle Scholar
    • 25 American Institute of Ultrasound in Medicine, Bioeffects Committee. Bioeffects considerations for the safety of diagnostic ultrasound. J Ultrasound Med. 1988; 7: S1–S38.MedlineGoogle Scholar
    • 26 Gould KL, Lipscomb K, Hamilton GW. Physiologic basis for assessing critical coronary stenosis: instantaneous flow response and regional distribution during coronary hyperemia as measures of coronary flow reserve. Am J Cardiol. 1974; 33: 87–94.CrossrefMedlineGoogle Scholar
    • 27 Ward-Smith AJ. Internal Fluid Flow: The Fluid Dynamics of Flow in Pipes and Ducts. New York, NY: Oxford Press; 1980:378–434.Google Scholar
    • 28 Siddiqi F, Blinc A, Braaten J, Francis CW. Ultrasound increases flow through fibrin gels. Thromb Haemost. 1995; 73: 495–498.CrossrefMedlineGoogle Scholar
    • 29 Siddiqi F, Odrljin TM, Fay PJ, Cox C, Francis CW. Binding of tissue-plasminogen activator to fibrin: effect of ultrasound. Blood. 1998; 91: 2019–2025.CrossrefMedlineGoogle Scholar
    • 30 Braaten JV, Goss RA, Francis CW. Ultrasound reversibly disaggregates fibrin fibers. Thromb Haemost. 1997; 78: 1063–1068.CrossrefMedlineGoogle Scholar
    • 31 Blinc A, Francis CW. Transport processes in fibrinolysis and fibrinolytic therapy. Thromb Haemost. 1996; 76: 481–491.CrossrefMedlineGoogle Scholar
    • 32 Spengos K, Behrens S, Daffertshofer M, Dempfle CE, Hennerici M. Acceleration of thrombolysis with ultrasound through the cranium in a flow model. Ultrasound Med Biol. 2000; 26: 889–895.CrossrefMedlineGoogle Scholar
    • 33 Alexandrov AV, Demchuk AM, Felberg RA, Christou I, Barber PA, Burgin WS, Malkoff M, Wojner AW, Grotta JC. High rate of complete recanalization and dramatic clinical recovery during tPA infusion when continuously monitored with 2-MHz transcranial Doppler monitoring. Stroke. 2000; 31: 610–614.CrossrefMedlineGoogle Scholar
    • 34 Ishibashi T, Akiyama M, Furuhata H, Tokudome S, Ogawa T, Abe T. Ultrasonic transmittance at skull: a fundamental experiment for transcranial ultrasound therapy.In: Program and abstracts of the Seventh Meeting of the Neurosonology Research Group of the World Federation of Neurology. Winston-Salem, NC, August 12–17, 1997.Abstract.Google Scholar

    eLetters(0)

    eLetters should relate to an article recently published in the journal and are not a forum for providing unpublished data. Comments are reviewed for appropriate use of tone and language. Comments are not peer-reviewed. Acceptable comments are posted to the journal website only. Comments are not published in an issue and are not indexed in PubMed. Comments should be no longer than 500 words and will only be posted online. References are limited to 10. Authors of the article cited in the comment will be invited to reply, as appropriate.