Multimodality Imaging of Lower Extremity Peripheral Arterial Disease: Current Role and Future Directions

Introduction

Patients with symptomatic peripheral arterial disease (PAD) affecting the lower extremities are initially evaluated with an ankle-brachial index (ABI) and segmental pressure measurements. An ABI<0.90 diagnoses significant lower extremity PAD (>50% stenosis) with a sensitivity and specificity of 79% and 96%, respectively.1 An ABI between 0.90 and 1.0 is considered borderline. The localization of the stenosis can be inferred by an abnormal decrease (>20 mm Hg) in segmental lower extremity pressures. In the presence of calcified atherosclerosis, arteries may become stiff and noncompressible, which results in a falsely elevated ABI (often >1.3). For patients with suspected PAD and a normal ABI at rest, it is valuable to obtain postexercise ABI measurements, which if <0.85 are consistent with PAD and is an independent predictor of mortality.2 Imaging is then needed to confirm the location and degree of stenosis before revascularization or if the diagnosis of PAD is uncertain.
Characterization of PAD can be performed with noninvasive angiography using computed tomography (CTA) or magnetic resonance angiography (MRA), as well as with duplex ultrasonography (US), depending on patient specific characteristics (Table 1). Advances in both CTA and MRA provide clinicians with the opportunity to obtain a high resolution, 3-dimensional (3-D) road map of the peripheral arterial tree in patients, particularly when planning revascularization strategies. Invasive digital subtraction angiography (DSA) has been the accepted standard for evaluation of lower extremity atherosclerosis. Although DSA is a robust technique for diagnosing significant arterial stenosis or obstruction, it provides a 2-D view of the vessels, which may underestimate the degree of stenosis for tortuous vessels.3 Furthermore, there are inherent risks with arterial access, ionizing radiation, and the use of iodinated contrast media (CM).4
Table 1. Comparison Between CTA, MRA, and Duplex US for Diagnosis of PAD
Imaging ModalityStrengthsLimitations
CTARapid acquisitionHeavily calcified vessels
High spatial resolution
In-stent restenosis visualization better than MRAIodinated contrast
Faster scan time than MRAIonizing radiation exposure
MRABoth gadolinium enhanced and noncontrast approaches availableIn-stent restenosis limited by metal artifact depending on stent type Risk of NSF in patients with advanced renal disease who receive gadolinium
Neither iodinated contrast nor ionizing radiation are used
Duplex USEasily accessibleLess accurate in aortoiliac region because of patient obesity or intestinal gas
Inexpensive
No radiationLimited sensitivity for multilevel stenosis
No contrastHeavily calcified vessels
CTA indicates computed tomography angiography; MRA, magnetic resonance angiography; US, ultrasonography; NSF, nephrogenic systemic fibrosis; and PAD, peripheral arterial disease.
This review aims to deliver a comprehensive, yet concise review of noninvasive imaging modalities in the diagnosis of lower extremity PAD (defined as involving the aorta, iliacs, and infrainguinal arteries). Furthermore, we will integrate the available evidence and society guidelines, to provide an informed clinical perspective regarding the role of the imaging modalities in patient care. Briefly, future directions in computed tomography (CT) and magnetic resonance imaging (MRI) will be discussed as they pertain to clinical vascular medicine.

Computed Tomography Angiography

Background

Initially, CTA was limited by the use of single slice scanners that imaged a fraction of the arterial tree at a given time. With the advent of multidetector row CT, images of inflow and runoff vessels became possible using a single acquisition and CM injection. Multidetector row CT has progressed rapidly from 4-detector row scanners to 64-detector rows or more, resulting in increased speed of scanning and greater longitudinal field of view while maintaining near-isotropic imaging voxels. Minimum voxel dimension of 0.4 to 0.6 mm in the z-axis are currently achievable with 64-detector row scanners, optimizing visualization of smaller, distal arteries.
Image interpretation is aided by the use of advanced postprocessing techniques. Volume rendered 3-D reconstruction of the arterial tree provides a global overview for the rapid identification of pathology. Maximum intensity projection (MIP) images provide similar views to traditional angiography and are useful for qualitatively assessing the degree of stenosis. The usefulness of both volume rendered and MIP are limited by obscuration of the vessels by bones. Thus, automated software techniques for bone removal have been implemented; however, even the best algorithms require some degree of manual correction. Either automated or interpreter generated centerlines in the vessel of interest can be placed to obtain curved planar reformations, which gives both longitudinal and cross-sectional views of the vessel, which are useful for quantitative measurements; however, this is done at the cost of distortion of nearby anatomic landmarks. Distortion-free detailed interrogation is performed with thin section multiplanar reformats, which can be oriented to any desired plane. Axial 2-D source data are used when other techniques are inconclusive.
Vessels with dense calcifications have high attenuation, which may lead to blooming artifact and resultant overestimation of stenosis. To minimize this, a window level of 200 Hounsfield units and width of 1000 Hounsfield units are recommended5 for image analysis. In the setting of heavily calcified vessels or metallic stents, image reconstruction using a sharp kernel (B46) can reduce blooming artifact, although this results in increased background image noise.6 The true lumen in calcified arterial vessels is often best analyzed using a combination of reconstruction techniques and source images. Additionally, dual-energy CTA postprocessing to subtract voxels containing calcium in the vessel wall is a promising technique for evaluating calcified vessels.

Technical Considerations for CTA Image Acquisition

Imaging of the lower extremity peripheral vasculature is performed from the aortic diaphragmatic hiatus through the toes, to evaluate the entire inflow and runoff (Figure 1). In the upper abdomen, visualization of the splanchnic and renal vasculature relies on good breath holding to prevent obscuration of the vessel by motion artifact. Respiration does not affect imaging beyond the pelvis.
Figure 1. Computed tomography angiography of aortoiliac occlusive disease. Bone segmentation was achieved using automated algorithms with manual correction to produce full-volume maximum intensity projection (MIP) image after bone removal (A) and 3-dimensional volume rendered (VR) image with bones included at a partial transparency (B). Aortic occlusion is annotated on the MIP image by the asterisk directly inferior to the renal arteries. Occlusion of the bilateral common iliac arteries and stents, as well as the external iliac arteries, is also shown on both images. Reconstitution of runoff flow occurs at the superficial femoral arteries bilaterally. Collateral pathways are easily visualized on the VR images. The Arc of Riolan (arrow) supplies the inferior mesenteric artery (IMA) territory from the superior mesenteric artery. The superior rectal artery (arrowhead) supplies the internal iliac arteries (black arrows) from the IMA. The internal iliac artery collateralizes with the circumflex femoral arteries supply the runoff vessels.
CTA requires the use of iodinated CM for visualization of the vasculature. Injection rates vary depending on protocol, but typically are in the range of 4 to 6 mL/s delivered intravenously by a dual channel power injector. The amount of iodinated media will vary depending upon the patient and protocol used, with ≈100 to 120 mL used for an abdominal CTA with runoffs.5 However, this dose could be reduced to 80 mL in patients with a small body mass index. A saline chaser injected at the completion of the CM media injection results in greater enhancement of the vasculature and a reduction in the CM dose.
To maximize the arterial enhancement during the examination, both CM transit time and speed of acquisition need to be considered. CM transit time is the time from the beginning of CM injection to the arrival of CM in the territory of interest, and is derived by either real-time enhancement monitoring (bolus tracking) during the injection of full dose of CM or by a separate smaller test bolus performed before the exam. Historically, with slower scanners, the fastest acquisition speed was chosen and the duration of CM injection was set to the scan time with scan initiation determined by CM transit time. Because of the rapid acquisition of modern scanners, this approach results in under opacification of the peripheral vasculature as a result of insufficient time to fill the vascular tree. Adding 3 to 5 seconds to the CM transit time, reducing the pitch to a moderate value (typically around 0.9) and increasing the CM volume lengthens the scan duration allowing for better filling of the vasculature. However, pathology such as large aneurysms and high-grade stenosis can delay the transit of CM to the distal vasculature, resulting in scanning before its arrival. A second late scan of the calf should be preplanned in the event of inadequate pedal opacification.
A new strategy put forth by Fleischmann7 is to fix the duration of scanning the lower extremity vascular to 40 seconds by adjusting the pitch. The volume and rate of CM injection is based on weight and delivered as a bi-phasic injection to provide sustained opacification of the arterial system.
Many patients with PAD undergo multiple imaging studies during their lifetime; thus, reducing exposition to radiation is paramount in this population. Reducing tube voltage results in better attenuation of CM because of the energy of the photons being closer to the k-edge of iodine, and reduction in effective dose to the patient. A decrease from 120 to 100 kVp results in a 34% reduction in radiation dose without affecting diagnostic image quality.8 More aggressive dose reduction from 100 to 80 kVp maintains diagnostic image quality with an additional 23% dose savings.9 However, in larger patients, lower kVp protocols are limited by noise. The newest current generation scanners offer an additional dose reduction of 50% while maintaining image quality because of advances in image processing.10 Image domain-based iterative reconstruction algorithms, as compared with filtered back projection image reconstruction of prior generations scanners, reduce noise and maintain image sharpness through modeling techniques made possible by improvement in computer processing power. Alternatively, with a similar radiation dose, iterative reconstruction can be used to increase image sharpness without affecting noise levels, which preliminary studies suggest may improve visualization of the lumen of small stents or calcified vessels.11,12

Magnetic Resonance Angiography

Currently, high performance MR scanners provide striking angiographic images without patient exposure to radiation or the complexity of removing overlying bone from 3-D reconstructed images. MRA can be performed using contrast-enhanced (CE-MRA) approach or noncontrast-enhanced, flow-sensitive techniques that take advantage of the difference in signal properties between static tissue and flowing blood. These noncontrast MRA techniques include time-of-flight (TOF) MRA, ECG-gated partial-Fourier fast spin echo (FSE), and steady-state free precession (SSFP). In addition, phase-contrast MRA shows arterial blood flow direction and velocity.

Contrast-Enhanced MRA

CE-MRA is the workhorse of clinical MRA because of the robustness of the technique (Figure 2). CE-MRA takes advantage of the increase in arterial signal after administration of intravenous paramagnetic gadolinium (Gd)-based contrast agents because of the shortening of T1 relaxation in blood. Fast T1-weighted gradient-echo sequences with low flip angles are used in combination with an intravenous bolus injection, allowing for rapid acquisition of 3-D volumes with high signal-to-noise ratio of vessel to background. Visualization of the vasculature is further enhanced by mask subtraction to remove the background signal, similar to DSA. MRA interpretation relies heavily on full-volume MIP images created from the subtracted datasets, which provide a global prospective. Detailed image analysis is done using subvolume MIPs and multiplanar reformats.
Figure 2. Multistation contrast-enhanced magnetic resonance angiography of the abdomen, pelvis, and lower extremities was performed using stepping table acquisition and bi-phasic contrast media injection protocol. Imaging reveals a patent aorto-biiliac surgical bypass graft (extent defined by arrowheads), with bilateral runoff disease consisting of tandem stenoses of the left superficial femoral artery (SFA), punctuated with a focal high-grade stenosis and collateral formation (white arrow) and tandem stenoses with long segment occlusion of the right SFA and popliteal artery with collaterals supplied by the profunda femoral artery (black arrows).
In MR, images are derived from Fourier transformation of radiofrequency signals that are acquired during an imaging sequence. K-space is the raw image data as represented in frequency-time domain and is built up in a sequential, step-wise fashion. The data located at the center of k-space contribute most to the contrast agent within an image. Thus, for CE-MRA, acquisition of the center of k-space needs to occur during peak vessel enhancement. The timing of the scan initiation is determined from a test bolus or using bolus-tracking methods, similar to CTA.
There are 7 Gd-based contrast agents currently approved by the Food and Drug Administration in the United States, and 2 additional agents used in the European Union (Table 2). They are paramagnetic and classified on the basis of protein binding, linear or cyclic structure, and ionicity.13 In patients with normal renal function, the half-life of Gd-based contrast agents are ≈1.5 hours. All Gd-based contrast agents have been associated with the rare development of nephrogenic systemic fibrosis; however, the majority of cases involve the linear agents: Gadopentetate dimeglumine (Magnevist), Gadodiamide (Omniscan), and Gadoversetamide (Optimark). The current recommendation is to avoid any Gd-based contrast agent in patients with a glomerular filtration rate <30 mL/min per 1.73 m2 given risk of nephrogenic systemic fibrosis.14
Table 2. GBCA Used in Contrast-Enhanced MRA
GBCAExcretionProtein BindingStructureCharge
Gadopentetate dimeglumine (Magnevist)KidneysNoneLinearIonic
Gadobenate disodium (Multihance)Primarily kidneys (small amount in feces)Weak bindingLinearIonic
Gadodiamide (Omniscan)KidneysNoneLinearNonionic
Gadoversetamide (Optimark)KidneysNoneLinearNonionic
Gadoteridol (Prohance)KidneysNoneCyclicNonionic
Gadoxetate disodium (Eovist)Kidney and feces (enters liver cells)NoneLinearIonic
Gadofosveset trisodium (Ablavar)KidneysStrong-reversible bindingLinearIonic
MRA indicates magnetic resonance angiography; and GBCA, gadolinium-based contrast agent.
The older Gd-based contrast agents, move quickly from the vasculature to the interstitial space, which results in limited time for imaging with the maximal spatial resolution. However, a new blood pool contrast agent, gadofosveset trisodium, was recently approved by the Food and Drug Administration for 3-D MRA.15 This agent reversibly binds to albumin, increasing T1 relaxivity in a steady state, which lasts for about an hour. In a small study of 27 patients16 with 334 arterial segments, imaging with 0.03 mmol/kg of gadofosveset trisodium resulted in 100% agreement between degree of stenosis graded with steady-state MRA and DSA.
Multistation MRA can be performed to image the aorta and peripheral runoff arteries using the same bolus, typically ≈0.2 mmol/kg body weight, in a bolus-chase technique, wherein the aortoiliac, femoropopliteal, and tibial regions are sequentially imaged. Multistation MRA benefits from a bi-phasic injection protocol with a high initial rate of injection that is reduced for the lower stations, resulting in increased arterial opacification of the distal station. The routine use of parallel imaging, requiring phased array peripheral coils, has reduced typical breath holds requirements for the aortoiliac station from 15 to 20 seconds or less without significant loss of image quality.
Despite advances in acquisition speed, venous contamination of the tibial station can be problematic, particularly in the setting of critical limb ischemia, where there is a smaller time difference between arterial and venous enhancement. There are several options for addressing venous contamination,17 including thigh compression cuffs or a semicircular foam pillow behind the knee to impede lower extremity venous return. Another option is to image the calf vessels first with a separate contrast bolus (which eliminates the possibility of venous contamination). However, imaging the tibial station first using time-resolved MRA is the approach most commonly used in practice. Time-resolved MRA uses k-space undersampling with view-sharing to acquire multiple 3-D MRA datasets rapidly every 1 to 2 seconds during the injection of a small amount of contrast agent. Ninety seconds of imaging from the start of injection is sufficient for the calf in most cases. This dataset is subtracted from a mask image and is presented as a cine of MIP images, replicating the appearance of DSA and thus allowing form separation of arterial and venous phases.
The use of higher field strength MRI, such as 3.0 Tesla offer a higher signal-to-noise ratio, potential for lower doses of Gd and improved spatial resolution compared with 1.5 Tesla. In a small study of MRA at 3 Tesla, there was excellent agreement between degree of stenosis seen with MRA and conventional angiography,18 and there was nearly isotropic submillimeter 3-D voxels.

Noncontrast-Enhanced MRA

Noncontrast-enhanced MRA (Figure 3) was initially described by Wedeen et al19 and further developed by Miyazaki et al.20 The first such technique was 2-D TOF, a gradient-echo technique that depends on flow-related enhancement. There are several challenges to 2-D TOF, including inadequate signal for vessels deep in the abdomen, flow-related artifacts seen in areas of stenosis, difficulty in obtaining perfectly perpendicular imaging planes, and a lengthy scan time. The diagnostic performance of CE-MRA is >2-D TOF.21 However, 2-D TOF is useful for evaluation of tibial and pedal arterial stenosis or occlusion, particularily if inadequately visualized by contrast angiography and is as accurate as DSA for these vessels.22
Figure 3. Noncontrast magnetic resonance angiography (MRA). A, Two-dimensional (2-D) time-of-flight (TOF) MRA is acquired as multiple 2-D slices ideally oriented perpendicular to the vessel. The imaging slice is initially nulled and non-nulled blood is allowed to enter the slice at which time readout is performed. A selective saturation band is placed inferior to the imaging slice to suppress venous flow. Because of the 2-D, slice-by-slice nature of the technique, movement occurring between slice acquisitions appears as step artifact on coronal maximum intensity projection reconstruction. Two-dimensional TOF is susceptible to motion because of the long imaging times, as seen in this patient. B, Three-dimensional fast spin echo at 3 Tesla was acquired in 3 stations, revealing excellent imaging characteristics in this normal volunteer.
There has been renewed interest in noncontrast-enhanced MRA given concerns of nephrogenic systemic fibrosis secondary to Gd in patients with advanced renal disease. There are several new techniques23: (1) ECG-gated partial-Fourier FSE, (2) 3-D FSE technique with sampling perfection with application of optimized contrasts using different flip angle evolution (SPACE),24 (3) balanced SSFP with arterial spin labeling, and (4) quiescent-interval single shot MRA using 2-D SSFP.25
ECG-gated 3-D FSE exploits the difference in arterial and venous flow velocities during the cardiac cycle. During diastole arteries have lower blood flow, resulting in high signal intensity on T2-W images, whereas in systole, the higher arterial blood flow leads to low signal intensity. Venous blood is bright throughout the cardiac cycle because of the relatively low resting flow. The MRA is then obtained by subtracting systolic from diastolic images. In a small study at 1.5 Tesla,26 ECG-gated 3-D FSE was compared against CE-MRA and found to have a sensitivity and specificity of 85% and 76%, respectively, with an overestimation of stenosis in 22%. The use of parallel imaging and higher field strength may improve acquisition time and image quality.
3-D FSE with SPACE24 uses a noncontrast, T1-W dark blood whole body technique with a high degree of reliability in normal volunteers. It has improved k-space sampling efficiency compared with traditional 3-D FSE. Additional studies are needed to evaluate the test characteristics of this technique in patients with PAD.
Balanced SSFP with arterial spin labeling results in bright blood images in a single acquisition by first using a nonselective tag pulse to invert the magnetization for the entire imaging plane.23 Then a spatially selective tagging pulse is applied upstream of the arteries of interest and the tagged blood flows into the desired imaging region. In vessels with slow blood flow, the time needed to replace the tagged blood in the imaging plane can approach the T1 recovery time of blood, whereby the tagging effect would be lost.
Quiescent-interval single shot MRA using a modified single shot 2-D balanced SSFP pulse sequence25 has been evaluated in a pilot study of normal volunteers and subjects with PAD and can accurately depict peripheral vascular stenosis compared with CE-MRA with a sensitivity and specificity of 87% and 95%, respectively. Quiescent-interval single shot MRA does not use subtractive nonenhanced MRA methods. The sequence requires little optimization for an individual patient, and the peripheral MRA can be performed in <10 minutes. A 2-center study compared CE-MRA and quiescent-interval single shot MRA in 53 patients with known or suspected PAD27 and found a sensitivity and specificity of 85% to 90% and 95% to 97%, respectively, for the detection of significant stenosis.
Phase-contrast MRI shows a change in intravascular signal as arterial blood flow velocity deviates from the expected velocity encoded into the pulse sequence. Phase-contrast MRI has a sensitivity and specificity >90% for detection of clinically relevant stenosis (>75%) compared with DSA.28 However, Phase-contrast MRI is typically used only as an adjunct for determining degree of stenosis because of the long image acquisition time.23

Duplex US

Duplex US uses the combination of gray scale (B-mode) for vessel morphology and color pulsed wave Doppler techniques. It is safe, inexpensive, and can provide functional information about vessel stenosis.4 The primary criteria of intra-arterial peak systolic velocity (PSV) and ratio of PSV between the site of stenosis and adjacent normal vessel are used to gauge the degree of stenosis (Figure 4). A PSV ratio of >2.5 suggests a stenosis of 50% to 74% and if the end-diastolic velocity is >60 cm/s, this suggests a tighter stenosis, in the range of 75% to 99%.29 However, the most commonly used criteria is PSV ratio >2.0, along with PSV>200 cm/s, and aliasing and spectral broadening seen with color Doppler to diagnose a stenosis >50%.30 The addition of color flow imaging improves the diagnostic performance of duplex US for aortoiliac and femoropopliteal arteries.31 If no color Doppler signal is visible in the vessel, arterial occlusion is suggested (Figure 5). The evaluation of infrapopliteal vessels although feasible by US is time consuming and may be technically challenging. Thus, a complete evaluation of the native lower extremity arterial system may not be feasible in many patients.
Figure 4. Duplex ultrasonography showing in-stent restenosis of a right superficial femoral artery (SFA) stent (for reference, the proximal SFA velocity is 96 cm/s, giving a peak systolic velocity [PSV] ratio of 278/96 cm/s, or 2.9). EDV indicates end-diastolic velocity; and RI, resistance index.
Figure 5. Duplex ultrasonography showing an occluded superficial femoral artery (SFA) stent without any evidence of flow on color or pulsed wave Doppler.
For evaluation with duplex US, patient body habitus may limit evaluation of the aortoiliac vessels because of abdominal girth or intestinal gas. The quality of a duplex US can vary depending on the examiner’s experience, and there is only moderate interobserver agreement for duplex evaluation of the aortoiliac and femoropopliteal areas.32 In the presence of sequential stenosis, US is less sensitive for detection of further downstream lesions.

Clinical Applications

There are several important considerations when choosing a noninvasive imaging modality to characterize lower extremity PAD: (1) imaging time, (2) need for evaluation of nearby anatomy, such as renovascular disease, (3) whether the indication is for native vessel disease or follow-up after a revascularization (bypass or stenting), and (4) patient specific factors, such as kidney disease, diabetes mellitus, or the presence of implantable metal devices. There are 3 sets of clinical guidelines which apply to the use of noninvasive imaging with lower extremity PAD: the European Society of Cardiology, American Heart Association/American College of Cardiology and Trans-Atlantic Inter-Society Consensus (TASC) II.
Rapid image acquisition is a strength of CTA. CTA studies can be performed with scan times <5 minutes compared with MRA (≈20–30 minutes) or duplex US (30–45 minutes). If both legs are evaluated by duplex US, the scan times are quite long (≈1 hour) and may hinder patient flow in the vascular laboratory.

Native Vessel Disease

The TASC II guidelines4 and European Society of Cardiology33 recommend obtaining a MRA, or CTA, or duplex US for imaging of native vessel disease, depending upon local availability, cost, and experience. American Heart Association/American College of Cardiology30 guidelines support the use of CTA and MRA in the diagnosis of the anatomic location and presence of significant stenosis in patients with native vessel lower extremity PAD. Imaging with either CTA or MRA is well suited for the diagnosis of patients with lower extremity PAD, who have suspected aortic aneurysm/dissection, renovascular disease, trauma, or acute limb ischemia.
When choosing an imaging modality for native vessel disease, clinicians should consider whether the patient has underlying renal insufficiency, diabetes mellitus, or implanted metal devices. For patients who have normal renal function and are not diabetic, initial evaluation with either CTA or MRA is reasonable based on their similar excellent diagnostic capabilities. Overall, duplex US is a less sensitive technique for imaging native vessel stenosis compared with CTA or MRA; however, the greatest limitation of duplex US is the time required for evaluation of 2 lower extremities. A brief review of the important recent trials supporting the recommendation for either CTA or MRA in the diagnosis of native vessel disease follows.
A recent systematic review of the diagnostic performance of CTA included 20 studies using DSA as the accepted standard.34 This analysis included studies which used 2 to 4 detector row CT scanners as well as the contemporary 16 to 64 detector rows. In this population of nearly 1000 patients, 68% had symptomatic PAD and sensitivity of CTA for stenosis >50% or occlusion was 95% (95% CI, 92%–97%) with a specificity of 96% (95% CI, 93%–97%). Regardless of the location, excellent test characteristics were seen in the aortoiliac (sensitivity 96%, specificity 98%), femoropopliteal (sensitivity 97%, specificity 94%), and tibial arteries (sensitivity 95%, specificity 91%). Traditionally, the diagnostic performance of CTA in tibial disease is lower compared with the aorta-iliac and femoral levels, particularly in the setting of heavily calcified vessels.
A more recent single-center study evaluated the diagnostic performance of 64-slice CTA compared with DSA35 for detection of stenosis >70% on a per segment basis. CTA had an accuracy of 98%, sensitivity of 99%, and specificity of 97%. Furthermore, this study compared the clinical decision making outcomes using the TASC II4 guidelines based on CTA and DSA findings, which were identical in all patients but one who developed critical limb ischemia following CTA.
A recent meta-analysis was performed of CE-MRA for determination of >50% stenosis or occlusion in patients with symptomatic PAD36 compared with DSA. The pooled sensitivity of MRA was excellent at 95% and specificity 96%. Another meta-analysis37 of CE-MRA for the diagnosis of PAD compared with DSA demonstrated multiple studies with sensitivities and specificities >95%. Revascularization plans based on MRA agreed with the independent DSA in 90% of cases.38 Furthermore, a strategy of preoperative planning with MRA instead of DSA was expected to result in a 31% cost savings, highlighting the important role of preoperative noninvasive imaging as a cost savings measure.
Duplex US for the detection of stenosis >50% or occlusion was evaluated in a meta-analysis of 16 studies,39 which found good sensitivity for aortoiliac lesions (86%) and femoropopliteal (80%) with an excellent specificity of 96% or greater for both. In a single-center study, 152 patients with symptomatic PAD were evaluated with duplex US, MRA, and DSA.40 The diagnostic accuracy for hemodynamically significant stenosis involving the aorta, iliac, and superficial femoral arteries (SFAs) was 89% for duplex US and 94% for MRA compared with DSA. On the basis of a meta-analysis, the pooled sensitivity of 88% for duplex US was significantly lower than MRA (99%) for the detection of PAD.41 However, their pooled specificities were similar (96% for MRA and 95% for duplex US). No significant difference was noted in the diagnosis of stenosis above or below the knee.39
Given that clinicians have an option of MRA, CTA, or duplex US for the noninvasive imaging of lower extremity PAD, it is important to evaluate the results from clinical trials comparing these modalities with regards to clinical outcomes. The Diagnostic Imaging of Peripheral Arterial Disease (DIPAD) trial evaluated clinical confidence in treatment plans and cost in >500 patients imaged with a duplex US, or MRA at 1.5 Tesla, or 16-detector row CTA.42 After adjusting for baseline variables, clinical confidence was higher and less additional imaging was required with the initial use of either MRA or CTA compared with duplex US. Importantly, the total cost for CTA were significantly lower compared with MRA or duplex US (because of the need for additional imaging in the US group). A randomized controlled trial compared CE-MRA and CTA for the initial diagnostic imaging as well as total diagnostic costs.43 No difference was found in the number of patients who underwent additional vascular imaging within 60 days of their initial study. The total diagnostic cost per patient was higher for MRA because of the baseline higher cost.

Prior Bypass Graft

Either duplex US, or CTA, or MRA can be used to evaluate for graft patency after lower extremity bypass surgery. The American Heart Association/American College of Cardiology recommends a surveillance program for postoperative infrainguinal vein bypass grafts with duplex US at 3, 6, 12, and 24 months after surgery (class I, level of evidence [LOE] A). Duplex US is ideally suited for postlower extremity arterial bypass surgery surveillance of femoral-popliteal or femoral-tibial/pedal venous grafts. Asymptomatic vein graft stenosis can result in acute thrombosis and ultimately graft failure if not detected early. In one study,44 vein grafts which were revised on the basis of stenosis findings with duplex US, had a 1-year patency rate of 90% compared with those grafts which were not revised and which had a patency rate of only 66%. However, duplex US is not well established for evaluating long-term patency of angioplasty or femoral-popliteal bypass with a synthetic conduit (class IIb, level of evidence [LOE] B).
When duplex US was compared with 4-row CTA, there was excellent agreement for the diagnosis of graft stenosis, aneurysmal changes, and arteriovenous fistulas.45 There was also no difference between duplex US and CTA when compared with DSA. Compared with duplex US, CTA may be superior for the detection of clinically significant stenosis in PAD bypass grafts.45

Stents

CTA is superior to MRA for evaluation of restenosis in metallic stents or stent grafts. There are varying types of metal vascular stents, which influence the ability of a MRA to determine the degree of in-stent restenosis reliably (Figure 6).17 The lumen within a steel stent may be completely obscured because of susceptibility artifact and radiofrequency shielding. Stents made with nitinol, cobalt, and platinum are less sensitive to these effects. Stents oriented with the long axis parallel to the B0 field of the scanner demonstrate better imaging characteristic compared with those oriented perpendicularly.
Figure 6. Magnetic resonance angiography showing a clearly patent right superficial femoral artery stent (double arrows). This is a nitinol stent which does not have the metal dropout artifact on magnetic resonance imaging.
Li et al46 used 64-detector row CTA to determine in-stent restenosis in 41 patients compared with DSA. Of the 81 stents, 76% were considered to be assessable based on image quality; motion or metal artifact influenced the interpretability of the other stents. For the group of evaluable stents, there was excellent sensitivity (95%) and specificity (96%) for detection of in-stent stenosis. When comparing the total group of in-stent stenosis seen on DSA with all CTA images (evaluable and nonevaluable stents), there was a lower sensitivity of 86% and specificity of 72% for CTA detection of in-stent stenosis. Beam hardening artifact from stents still limits detection of in-stent restenosis with CTA. Future in vivo studies will be required to determine whether iterative reconstruction can improve visualization.
Duplex US can be used to determine areas of in-stent restenosis, particularly within the iliac, SFA, or popliteal arteries (Figures 4 and 5). To identify >80% stenosis of the SFA compared with DSA,47 a PSV>275 cm/s and velocity ratio (comparing the velocity within the stent to a diseased free area vessel proximal to the stent) of >3.5 has excellent specificity (94%) but only moderate sensitivity (74%). The limited sensitivity of duplex US for detection of significant stenosis results in CTA being the diagnostic modality of choice for in-stent restenosis unless there are concerns regarding the potential for renal failure with the iodinated CM.

Special Populations

Patients with Diabetes

Patients with diabetes mellitus often have heavily calcified vessels, which presents a diagnostic challenge in terms of using duplex US or CTA. With duplex US, heavily calcified vessels may result in an obscured arterial lumen and inability to diagnose the degree of PAD reliably. CE-MRA is not subject to image artifact generated by heavily calcified vessels (Figure 7). Interobserver agreement for degree of stenosis or occlusion is higher with CE-MRA than CTA, in the presence of arterial calcifications.48 From our perspective, for patients with diabetes, MRA should be considered the test of choice because of calcified vessels and concomitant renal insufficiency.
Figure 7. Vascular calcifications in a patient with diabetes mellitus. A, Coronal thin section maximum intensity projection (MIP) of a computed tomography angiography of the tibial peroneal runoff vessels demonstrating how dense calcifications in this patient population obscures analysis of the lumen. B, Contrast-enhanced magnetic resonance angiography is insensitive to vascular calcification as well as bone, simplifying image interpretation as demonstrated in this full-volume MIP from the same patient as in A.

Renal Failure

As discussed previously, because of concerns regarding nephrogenic systemic fibrosis, the current recommendation is to avoid gadolinium-based contrast agents14 in patients with advanced renal failure with an estimated glomerular filtration rate <30 mL/min per 1.73 m2 unless the information provided by the CE-MRI is not obtainable by another means. Patients with any degree of renal insufficiency can be safely imaged with duplex US and noncontrast MRA. With CTA, iodinated CM has an associated risk of contrast-induced nephropathy defined as an increase in serum creatinine from baseline by >25% or >0.5 mg/dL within 3 days of contrast agent administration without other cause. Those at highest risk for contrast-induced nephropathy have baseline renal insufficiency, diabetes mellitus, concurrent nephrotoxic drug use, hypertension, congestive heart failure, and dehydration. Large doses of CM and high-osmolar CM may further increase the risk of contrast-induced nephropathy.49 The greatest risk seems to be in patients with pre-existing renal insufficiency, particularly those with diabetes mellitus, regardless of the osmolality of the contrast agent used.50 It is important to ensure adequate hydration and to minimize the total volume of CM used. In addition, patients with allergic reactions to iodinated CM need premedication with steroids and antihistamine before receiving it.

Implanted Metal Devices

For patients with implanted metal devices, it is important to determine compatibility with exposure to magnetic fields. Device information cards may have manufacturer information regarding safety testing with MRI. Publications are available which consolidate the available information regarding metal devices and MRI safety.51 Newer pacemakers are being developed that are MRI-compatible. Patients who are pacemaker-dependent or have certain types of brain aneurysm clips, neurostimulator devices, or retained ocular ferromagnetic foreign bodies are best imaged with another modality because of potential risk. In addition, even if an implanted medical device is considered compatible with the MRI environment, there may be significant susceptibility artifact generated during the scan.

Clinical Application Summary

Given the options available for multimodality imaging of lower extremity PAD, we present an overview on the basis of our perspective and review of the literature (Figure 8).
Figure 8. An algorithm for choosing the appropriate imaging modality for a given clinical scenario. PAD indicates peripheral arterial disease; CTA, computed tomography angiography; US, ultrasonography; CE-MRA, contrast-enhanced magnetic resonance angiography; NCE, non-contrast enhanced; and GFR, glomerular filtration rate.

Future Directions

Atherosclerotic Plaque Characterization

The degree of atherosclerotic plaque can be inferred by the amount and degree of stenosis seen on noninvasive imaging. However, recent studies, especially with MRI, have sought to understand atherosclerotic plaque burden and component characterization of the lower extremities to predict plaque stability and response to drug therapy. The total plaque volume in the SFA can be reliably measured by MRI52 and was used in a clinical trial of lipid lowering therapy.53 Atherosclerotic plaque composition is linked with vulnerability and progression, therefore multicontrast MRI showing calcification and lipid rich necrotic core54 may be useful in clinical trials of PAD.
Atherosclerotic plaque composition in the SFA can also be evaluated with 18F-fluoro-deoxy-glucose-positron emission tomography/CT (FDG-PET/CT),55 which identifies areas of inflammation (PET) and calcification (CT) and has excellent reproducibility in clinical trials. Complementary MRI and FDG-PET/CT have been used to evaluate atherosclerosis in the SFA.56 FDG-PET/CT shows a higher vascular FDG uptake within lipid rich plaque visualized by MRI.56

Calf Muscle Metabolism and Perfusion

Current research using CT, MRI, and US is aimed at simultaneous evaluation of lower extremity anatomy as well as physiology to better understand which treatment options are best for a given patient. A recent study using FDG-PET in patients with claudication found evidence of impaired calf muscle glucose uptake,57 which did not correlate with baseline blood glucose levels. Another method for noninvasive evaluation of skeletal muscle metabolism in PAD is 31-phosphorus magnetic resonance spectroscopy. PAD patients have delayed recovery of phosphocreatine energy stores after exercise,58 which does not correlate well with MRI measures of calf muscle perfusion, suggesting uncoupling of perfusion and energetics and indicating an intrinsic mitochondrial defect in calf skeletal muscle in PAD.59 Magnetic resonance spectroscopy used for evaluation of patients with PAD is currently limited to the research arena.
Calf muscle perfusion can be measured with CE-MRI first pass perfusion with Gd at peak exercise,60 with noncontrast-enhanced MRI using arterial spin labeling after exercise61 or hyperemia related to thigh cuff occlusion and release.62 Contrast US with microbubbles has also been used to evaluate lower extremity arterial perfusion reserve.63 These techniques may prove useful in development of novel angiogenic therapies for patients with PAD.

Conclusions

Noninvasive imaging for lower extremity PAD is indicated to identify the location and severity of arterial stenosis when considering possible intervention. The major guidelines (TASC II, European Society of Cardiology, and American College of Cardiology/American Heart Association) support the use of duplex US, CTA, or MRA depending upon patient specific factors, local expertise, and safety profile. Promising research studies are providing new insights into using noninvasive imaging for atherosclerotic plaque characterization and lower extremity physiology assessment.

References

1.
Lijmer JG, Hunink MG, van den Dungen JJ, Loonstra J, Smit AJ. ROC analysis of noninvasive tests for peripheral arterial disease. Ultrasound Med Biol. 1996;22:391–398.
2.
Sheikh MA, Bhatt DL, Li J, Lin S, Bartholomew JR. Usefulness of postexercise ankle-brachial index to predict all-cause mortality. Am J Cardiol. 2011;107:778–782.
3.
Ota H, Takase K, Rikimaru H, Tsuboi M, Yamada T, Sato A, Higano S, Ishibashi T, Takahashi S. Quantitative vascular measurements in arterial occlusive disease. Radiographics. 2005;25:1141–1158.
4.
Norgren L, Hiatt WR, Dormandy JA, Nehler MR, Harris KA, Fowkes FG; TASC II Working Group. Inter-Society Consensus for the Management of Peripheral Arterial Disease (TASC II). J Vasc Surg. 2007;45 Suppl S:S5–67.
5.
Kock MC, Dijkshoorn ML, Pattynama PM, Myriam Hunink MG. Multi-detector row computed tomography angiography of peripheral arterial disease. Eur Radiol. 2007;17:3208–3222.
6.
Suzuki S, Furui S, Kaminaga T, Yamauchi T, Kuwahara S, Yokoyama N, Suzuki M, Isshiki T. Evaluation of coronary stents in vitro with CT angiography: effect of stent diameter, convolution kernel, and vessel orientation to the z-axis. Circ J. 2005;69:1124–1131.
7.
Fleischmann D. CT angiography: injection and acquisition technique. Radiol Clin North Am. 2010;48:237–47, vii.
8.
Wintersperger B, Jakobs T, Herzog P, Schaller S, Nikolaou K, Suess C, Weber C, Reiser M, Becker C. Aorto-iliac multidetector-row CT angiography with low kV settings: improved vessel enhancement and simultaneous reduction of radiation dose. Eur Radiol. 2005;15:334–341.
9.
Schindera ST, Graca P, Patak MA, Abderhalden S, von Allmen G, Vock P, Szucs-Farkas Z. Thoracoabdominal-aortoiliac multidetector-row CT angiography at 80 and 100 kVp: assessment of image quality and radiation dose. Invest Radiol. 2009;44:650–655.
10.
Winklehner A, Karlo C, Puippe G, Schmidt B, Flohr T, Goetti R, Pfammatter T, Frauenfelder T, Alkadhi H. Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential. Eur Radiol. 2011;21:2521–2526.
11.
Min JK, Swaminathan RV, Vass M, Gallagher S, Weinsaft JW. High-definition multidetector computed tomography for evaluation of coronary artery stents: comparison to standard-definition 64-detector row computed tomography. J Cardiovasc Comput Tomogr. 2009;3:246–251.
12.
Renker M, Nance JW, Schoepf UJ, O’Brien TX, Zwerner PL, Meyer M, Kerl JM, Bauer RW, Fink C, Vogl TJ, Henzler T. Evaluation of heavily calcified vessels with coronary CT angiography: comparison of iterative and filtered back projection image reconstruction. Radiology. 2011;260:390–399.
13.
Morcos SK. Extracellular gadolinium contrast agents: differences in stability. Eur J Radiol. 2008;66:175–179.
14.
Kribben A, Witzke O, Hillen U, Barkhausen J, Daul AE, Erbel R. Nephrogenic systemic fibrosis: pathogenesis, diagnosis, and therapy. J Am Coll Cardiol. 2009;53:1621–1628.
15.
Goyen M, Shamsi K, Schoenberg SO. Vasovist-enhanced MR angiography. Eur Radiol. 2006;16 Suppl 2:B9–14.
16.
Hadizadeh DR, Gieseke J, Lohmaier SH, Wilhelm K, Boschewitz J, Verrel F, Schild HH, Willinek WA. Peripheral MR angiography with blood pool contrast agent: prospective intraindividual comparative study of high-spatial-resolution steady-state MR angiography versus standard-resolution first-pass MR angiography and DSA. Radiology. 2008;249:701–711.
17.
Roditi G, Kusumawidjaja D. Magnetic resonance angiography and computed tomography angiography for peripheral arterial disease. Imaging. 2009;21:85–108.
18.
Nael K, Krishnam M, Nael A, Ton A, Ruehm SG, Finn JP. Peripheral contrast-enhanced MR angiography at 3.0T, improved spatial resolution and low dose contrast: initial clinical experience. Eur Radiol. 2008;18:2893–2900.
19.
Wedeen VJ, Meuli RA, Edelman RR, Geller SC, Frank LR, Brady TJ, Rosen BR. Projective imaging of pulsatile flow with magnetic resonance. Science. 1985;230:946–948.
20.
Miyazaki M, Sugiura S, Tateishi F, Wada H, Kassai Y, Abe H. Non-contrast-enhanced MR angiography using 3D ECG-synchronized half-Fourier fast spin echo. J Magn Reson Imaging. 2000;12:776–783.
21.
Koelemay MJ, Lijmer JG, Stoker J, Legemate DA, Bossuyt PM. Magnetic resonance angiography for the evaluation of lower extremity arterial disease: a meta-analysis. JAMA. 2001;285:1338–1345.
22.
McCauley TR, Monib A, Dickey KW, Clemett J, Meier GH, Egglin TK, Gusberg RJ, Rosenblatt M, Pollak JS. Peripheral vascular occlusive disease: accuracy and reliability of time-of-flight MR angiography. Radiology. 1994;192:351–357.
23.
Miyazaki M, Lee VS. Nonenhanced MR angiography. Radiology. 2008;248:20–43.
24.
Mihai G, Chung YC, Merchant A, Simonetti OP, Rajagopalan S. T1-weighted-SPACE dark blood whole body magnetic resonance angiography (DB-WBMRA): initial experience. J Magn Reson Imaging. 2010;31:502–509.
25.
Edelman RR, Sheehan JJ, Dunkle E, Schindler N, Carr J, Koktzoglou I. Quiescent-interval single-shot unenhanced magnetic resonance angiography of peripheral vascular disease: Technical considerations and clinical feasibility. Magn Reson Med. 2010;63:951–958.
26.
Lim RP, Hecht EM, Xu J, Babb JS, Oesingmann N, Wong S, Muhs BE, Gagne P, Lee VS. 3D nongadolinium-enhanced ECG-gated MRA of the distal lower extremities: preliminary clinical experience. J Magn Reson Imaging. 2008;28:181–189.
27.
Hodnett PA, Koktzoglou I, Davarpanah AH, Scanlon TG, Collins JD, Sheehan JJ, Dunkle EE, Gupta N, Carr JC, Edelman RR. Evaluation of peripheral arterial disease with nonenhanced quiescent-interval single-shot MR angiography. Radiology. 2011;260:282–293.
28.
Reimer P, Boos M. Phase-contrast MR angiography of peripheral arteries: technique and clinical application. Eur Radiol. 1999;9:122–127.
29.
de Smet AA, Ermers EJ, Kitslaar PJ. Duplex velocity characteristics of aortoiliac stenoses. J Vasc Surg. 1996;23:628–636.
30.
Rooke TW, Hirsch AT, Misra S, Sidawy AN, Beckman JA, Findeiss LK, Golzarian J, Gornik HL, Halperin JL, Jaff MR, Moneta GL, Olin JW, Stanley JC, White CJ, White JV, Zierler RE; Society for Cardiovascular Angiography and Interventions; Society of Interventional Radiology; Society for Vascular Medicine; Society for Vascular Surgery. 2011 ACCF/AHA Focused Update of the Guideline for the Management of Patients With Peripheral Artery Disease (updating the 2005 guideline): a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2011;58:2020–2045.
31.
de Vries SO, Hunink MG, Polak JF. Summary receiver operating characteristic curves as a technique for meta-analysis of the diagnostic performance of duplex ultrasonography in peripheral arterial disease. Acad Radiol. 1996;3:361–369.
32.
Ubbink DT, Fidler M, Legemate DA. Interobserver variability in aortoiliac and femoropopliteal duplex scanning. J Vasc Surg. 2001;33:540–545.
33.
Tendera M, Aboyans V, Bartelink ML, Baumgartner I, Clément D, Collet JP, Cremonesi A, De Carlo M, Erbel R, Fowkes FG, Heras M, Kownator S, Minar E, Ostergren J, Poldermans D, Riambau V, Roffi M, Röther J, Sievert H, van Sambeek M, Zeller T; European Stroke Organisation; ESC Committee for Practice Guidelines. ESC Guidelines on the diagnosis and treatment of peripheral artery diseases: Document covering atherosclerotic disease of extracranial carotid and vertebral, mesenteric, renal, upper and lower extremity arteries: the Task Force on the Diagnosis and Treatment of Peripheral Artery Diseases of the European Society of Cardiology (ESC). Eur Heart J. 2011;32:2851–2906.
34.
Met R, Bipat S, Legemate DA, Reekers JA, Koelemay MJ. Diagnostic performance of computed tomography angiography in peripheral arterial disease: a systematic review and meta-analysis. JAMA. 2009;301:415–424.
35.
Napoli A, Anzidei M, Zaccagna F, Cavallo Marincola B, Zini C, Brachetti G, Cartocci G, Fanelli F, Catalano C, Passariello R. Peripheral arterial occlusive disease: diagnostic performance and effect on therapeutic management of 64-section CT angiography. Radiology. 2011;261:976–986.
36.
Menke J, Larsen J. Meta-analysis: Accuracy of contrast-enhanced magnetic resonance angiography for assessing steno-occlusions in peripheral arterial disease. Ann Intern Med. 2010;153:325–334.
37.
Dellegrottaglie S, Sanz J, Macaluso F, Einstein AJ, Raman S, Simonetti OP, Rajagopalan S. Technology Insight: magnetic resonance angiography for the evaluation of patients with peripheral artery disease. Nat Clin Pract Cardiovasc Med. 2007;4:677–687.
38.
Hoch JR, Tullis MJ, Kennell TW, McDermott J, Acher CW, Turnipseed WD. Use of magnetic resonance angiography for the preoperative evaluation of patients with infrainguinal arterial occlusive disease. J Vasc Surg. 1996;23:792–800; discussion 801.
39.
Koelemay MJ, den Hartog D, Prins MH, Kromhout JG, Legemate DA, Jacobs MJ. Diagnosis of arterial disease of the lower extremities with duplex ultrasonography. Br J Surg. 1996;83:404–409.
40.
Leiner T, Kessels AG, Nelemans PJ, Vasbinder GB, de Haan MW, Kitslaar PE, Ho KY, Tordoir JH, van Engelshoven JM. Peripheral arterial disease: comparison of color duplex US and contrast-enhanced MR angiography for diagnosis. Radiology. 2005;235:699–708.
41.
Visser K, Hunink MG. Peripheral arterial disease: gadolinium-enhanced MR angiography versus color-guided duplex US–a meta-analysis. Radiology. 2000;216:67–77.
42.
Ouwendijk R, de Vries M, Stijnen T, Pattynama PM, van Sambeek MR, Buth J, Tielbeek AV, van der Vliet DA, SchutzeKool LJ, Kitslaar PJ, de Haan MW, van Engelshoven JM, Hunink MG; Program for the Assessment of Radiological Technology. Multicenter randomized controlled trial of the costs and effects of noninvasive diagnostic imaging in patients with peripheral arterial disease: the DIPAD trial. AJR Am J Roentgenol. 2008;190:1349–1357.
43.
Ouwendijk R, de Vries M, Pattynama PM, van Sambeek MR, de Haan MW, Stijnen T, van Engelshoven JM, Hunink MG. Imaging peripheral arterial disease: a randomized controlled trial comparing contrast-enhanced MR angiography and multi-detector row CT angiography. Radiology. 2005;236:1094–1103.
44.
Mattos MA, van Bemmelen PS, Hodgson KJ, Ramsey DE, Barkmeier LD, Sumner DS. Does correction of stenoses identified with color duplex scanning improve infrainguinal graft patency? J Vasc Surg. 1993;17:54–64; discussion 64.
45.
Willmann JK, Mayer D, Banyai M, Desbiolles LM, Verdun FR, Seifert B, Marincek B, Weishaupt D. Evaluation of peripheral arterial bypass grafts with multi-detector row CT angiography: comparison with duplex US and digital subtraction angiography. Radiology. 2003;229:465–474.
46.
Li XM, Li YH, Tian JM, Xiao Y, Lu JP, Jing ZP, Sheng J, Edwin A, Wu FH. Evaluation of peripheral artery stent with 64-slice multi-detector row CT angiography: prospective comparison with digital subtraction angiography. Eur J Radiol. 2010;75:98–103.
47.
Baril DT, Rhee RY, Kim J, Makaroun MS, Chaer RA, Marone LK. Duplex criteria for determination of in-stent stenosis after angioplasty and stenting of the superficial femoral artery. J Vasc Surg. 2009;49:133–8; discussion 139.
48.
Ouwendijk R, Kock MC, Visser K, Pattynama PM, de Haan MW, Hunink MG. Interobserver agreement for the interpretation of contrast-enhanced 3D MR angiography and MDCT angiography in peripheral arterial disease. AJR Am J Roentgenol. 2005;185:1261–1267.
49.
Rudnick MR, Goldfarb S, Wexler L, Ludbrook PA, Murphy MJ, Halpern EF, Hill JA, Winniford M, Cohen MB, VanFossen DB. Nephrotoxicity of ionic and nonionic contrast media in 1196 patients: a randomized trial. The Iohexol Cooperative Study. Kidney Int. 1995;47:254–261.
50.
Spinazzi A, Pozzi Mucelli R. [Administration of iodinated contrast in patients with pre-existing renal failure: a review]. Radiol Med. 2004;107:88–97.
51.
Sawyer-Glover AM, Shellock FG. Pre-MRI procedure screening: recommendations and safety considerations for biomedical implants and devices. J Magn Reson Imaging. 2000;12:92–106.
52.
Isbell DC, Meyer CH, Rogers WJ, Epstein FH, DiMaria JM, Harthun NL, Wang H, Kramer CM. Reproducibility and reliability of atherosclerotic plaque volume measurements in peripheral arterial disease with cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2007;9:71–76.
53.
West AM, Anderson JD, Meyer CH, Epstein FH, Wang H, Hagspiel KD, Berr SS, Harthun NL, DiMaria JM, Hunter JR, Christopher JM, Chew JD, Winberry GB, Kramer CM. The effect of ezetimibe on peripheral arterial atherosclerosis depends upon statin use at baseline. Atherosclerosis. 2011;218:156–162.
54.
Li F, McDermott MM, Li D, Carroll TJ, Hippe DS, Kramer CM, Fan Z, Zhao X, Hatsukami TS, Chu B, Wang J, Yuan C. The association of lesion eccentricity with plaque morphology and components in the superficial femoral artery: a high-spatial-resolution, multi-contrast weighted CMR study. J Cardiovasc Magn Reson. 2010;12:37.
55.
Rudd JH, Myers KS, Bansilal S, Machac J, Pinto CA, Tong C, Rafique A, Hargeaves R, Farkouh M, Fuster V, Fayad ZA. Atherosclerosis inflammation imaging with 18F-FDG PET: carotid, iliac, and femoral uptake reproducibility, quantification methods, and recommendations. J Nucl Med. 2008;49:871–878.
56.
Silvera SS, Aidi HE, Rudd JH, Mani V, Yang L, Farkouh M, Fuster V, Fayad ZA. Multimodality imaging of atherosclerotic plaque activity and composition using FDG-PET/CT and MRI in carotid and femoral arteries. Atherosclerosis. 2009;207:139–143.
57.
Pande RL, Park MA, Perlstein TS, Desai AS, Doyle J, Navarrete N, Copeland-Halperin RS, Redline W, Di Carli MF, Creager MA. Impaired skeletal muscle glucose uptake by [18F]fluorodeoxyglucose-positron emission tomography in patients with peripheral artery disease and intermittent claudication. Arterioscler Thromb Vasc Biol. 2011;31:190–196.
58.
Isbell DC, Berr SS, Toledano AY, Epstein FH, Meyer CH, Rogers WJ, Harthun NL, Hagspiel KD, Weltman A, Kramer CM. Delayed calf muscle phosphocreatine recovery after exercise identifies peripheral arterial disease. J Am Coll Cardiol. 2006;47:2289–2295.
59.
Anderson JD, Epstein FH, Meyer CH, Hagspiel KD, Wang H, Berr SS, Harthun NL, Weltman A, Dimaria JM, West AM, Kramer CM. Multifactorial determinants of functional capacity in peripheral arterial disease: uncoupling of calf muscle perfusion and metabolism. J Am Coll Cardiol. 2009;54:628–635.
60.
Isbell DC, Epstein FH, Zhong X, DiMaria JM, Berr SS, Meyer CH, Rogers WJ, Harthun NL, Hagspiel KD, Weltman A, Kramer CM. Calf muscle perfusion at peak exercise in peripheral arterial disease: measurement by first-pass contrast-enhanced magnetic resonance imaging. J Magn Reson Imaging. 2007;25:1013–1020.
61.
West AM, Meyer CH, Epstein FH, Jiji R, Hunter JR, DiMaria JM, Christopher JM, Kramer CM. Arterial spin labeling MRI reproducibly measures peak-exercise calf muscle perfusion in healthy volunteers and patients with peripheral vascular disease. JACC Cardiovasc Imag. In press.
62.
Wu WC, Mohler E, Ratcliffe SJ, Wehrli FW, Detre JA, Floyd TF. Skeletal muscle microvascular flow in progressive peripheral artery disease: assessment with continuous arterial spin-labeling perfusion magnetic resonance imaging. J Am Coll Cardiol. 2009;53:2372–2377.
63.
Lindner JR, Womack L, Barrett EJ, Weltman J, Price W, Harthun NL, Kaul S, Patrie JT. Limb stress-rest perfusion imaging with contrast ultrasound for the assessment of peripheral arterial disease severity. JACC Cardiovasc Imaging. 2008;1:343–350.

eLetters(0)

eLetters should relate to an article recently published in the journal and are not a forum for providing unpublished data. Comments are reviewed for appropriate use of tone and language. Comments are not peer-reviewed. Acceptable comments are posted to the journal website only. Comments are not published in an issue and are not indexed in PubMed. Comments should be no longer than 500 words and will only be posted online. References are limited to 10. Authors of the article cited in the comment will be invited to reply, as appropriate.

Comments and feedback on AHA/ASA Scientific Statements and Guidelines should be directed to the AHA/ASA Manuscript Oversight Committee via its Correspondence page.

Information & Authors

Information

Published In

Go to Circulation: Cardiovascular Imaging
Go to Circulation: Cardiovascular Imaging
Circulation: Cardiovascular Imaging
Pages: 797 - 807
PubMed: 23169982

History

Received: 31 January 2012
Accepted: 11 September 2012
Published online: 1 November 2012
Published in print: November 2012

Permissions

Request permissions for this article.

Keywords

  1. peripheral arterial disease
  2. computed tomographic angiography
  3. magnetic resonance angiography
  4. duplex ultrasonography

Subjects

Authors

Affiliations

Amy W. Pollak, MD
From the Departments of Medicine (A.W.P., P.T.N., C.M.K) and Radiology (P.T.N., C.M.K.), and the Cardiovascular Imaging Center, University of Virginia Health System, Charlottesville, VA.
Patrick T. Norton, MD
From the Departments of Medicine (A.W.P., P.T.N., C.M.K) and Radiology (P.T.N., C.M.K.), and the Cardiovascular Imaging Center, University of Virginia Health System, Charlottesville, VA.
Christopher M. Kramer, MD
From the Departments of Medicine (A.W.P., P.T.N., C.M.K) and Radiology (P.T.N., C.M.K.), and the Cardiovascular Imaging Center, University of Virginia Health System, Charlottesville, VA.

Notes

Correspondence to Christopher M. Kramer, MD, University of Virginia Health System, Departments of Medicine and Radiology, Lee St, Box 800170, Charlottesville, VA 22908. E-mail [email protected]

Disclosures

Dr Kramer receives research support from Siemens Healthcare. The other authors have no conflicts to report.

Sources of Funding

Supported by National Heart Lung Blood Institute R01 HL075792 (C.M.K.).

Metrics & Citations

Metrics

Citations

Download Citations

If you have the appropriate software installed, you can download article citation data to the citation manager of your choice. Select your manager software from the list below and click Download.

  1. Cost-effectiveness of diagnostic imaging modalities in symptomatic patients with lower limb peripheral arterial disease: discrete event simulation model, Frontiers in Public Health, 12, (2024).https://doi.org/10.3389/fpubh.2024.1367447
    Crossref
  2. Assessment of arterial blood flow in the microvasculature of the lower extremities in patients with chronic ischemia using a single-photon emission computed tomography (SPECT), Regional blood circulation and microcirculation, 23, 1, (37-43), (2024).https://doi.org/10.24884/1682-6655-2024-23-1-37-43
    Crossref
  3. Leveraging the Cardiovascular Team in Peripheral Artery Disease Diagnosis: A Call to Action, Journal of Multidisciplinary Healthcare, Volume 17, (2903-2910), (2024).https://doi.org/10.2147/JMDH.S466345
    Crossref
  4. Clinical physiology: the crucial role of MRI in evaluation of peripheral artery disease, American Journal of Physiology-Heart and Circulatory Physiology, 326, 5, (H1304-H1323), (2024).https://doi.org/10.1152/ajpheart.00533.2023
    Crossref
  5. Design and development of the DE-SPECT system: a clinical SPECT system for broadband multi-isotope imaging of peripheral vascular disease, Physics in Medicine & Biology, 69, 12, (125016), (2024).https://doi.org/10.1088/1361-6560/ad5266
    Crossref
  6. Facing the challenges of peripheral arterial disease in the era of emerging technologies, JVS-Vascular Insights, 2, (100095), (2024).https://doi.org/10.1016/j.jvsvi.2024.100095
    Crossref
  7. Multi-Modality Imaging of Atheromatous Plaques in Peripheral Arterial Disease: Integrating Molecular and Imaging Markers, International Journal of Molecular Sciences, 24, 13, (11123), (2023).https://doi.org/10.3390/ijms241311123
    Crossref
  8. Automatic force-controlled 3D photoacoustic system for human peripheral vascular imaging, Biomedical Optics Express, 14, 2, (987), (2023).https://doi.org/10.1364/BOE.481163
    Crossref
  9. Can Sublingual Nitrate Improve Visualization of Lower Limb Arteries on Computed Tomography Angiography?, Journal of Computer Assisted Tomography, 47, 4, (576-582), (2023).https://doi.org/10.1097/RCT.0000000000001446
    Crossref
  10. Musculoskeletal manifestations of COVID-19, Skeletal Radiology, 53, 10, (2009-2022), (2023).https://doi.org/10.1007/s00256-023-04549-4
    Crossref
  11. See more
Loading...

View Options

View options

PDF and All Supplements

Download PDF and All Supplements

PDF/EPUB

View PDF/EPUB
Login options

Check if you have access through your login credentials or your institution to get full access on this article.

Personal login Institutional Login
Purchase Options

Purchase this article to access the full text.

Purchase access to this article for 24 hours

Multimodality Imaging of Lower Extremity Peripheral Arterial Disease
Circulation: Cardiovascular Imaging
  • Vol. 5
  • No. 6

Purchase access to this journal for 24 hours

Circulation: Cardiovascular Imaging
  • Vol. 5
  • No. 6
Restore your content access

Enter your email address to restore your content access:

Note: This functionality works only for purchases done as a guest. If you already have an account, log in to access the content to which you are entitled.

Figures

Tables

Media

Share

Share

Share article link

Share

Comment Response